Abstract
Background Polysomnography (PSG) is the gold standard for diagnosing and monitoring sleep disorders, however, it is time-consuming and costly, as the application of the equipment can only be done by trained sleep technicians and the test must be conducted within a sleep laboratory. In this study we assessed the performance of the first wireless patch-based PSG system, the Onera Sleep Test System (STS), which can be applied by the patient and performed outside of the sleep laboratory in settings such as the home. To achieve this, sleep stage and physiological data from the Onera STS were compared to gold standard in-lab PSG.
Materials and methods The recordings were unsupervised to simulate a home-use environment. Epoch-by-epoch agreement was assessed by calculating sensitivity, specificity, accuracy, and Cohen’s kappa coefficient. Pearson’s correlation coefficients were calculated for multiple sleep parameters to measure the level of entire night agreement.
Results Substantial agreement with a Cohens kappa of 0.69 across all sleep stages was determined, which reached 0.81 when Stage N1 was removed from the analysis. A high accuracy, specificity, and sensitivity were found for wake N2, N3 and REM. Although specificity (95.25%) and accuracy (89.62%) were high for N1, sensitivity was low (27.19%). Sleep parameters calculated by sleep stage transitions, apnea hypopnea index and oxygen desaturation index, showed strong correlations.
Conclusions The Onera STS provides comparable clinical information to traditional PSG. Moreover, the application time was reduced by 77% which reduces the overall costs of PSG. These results open the possibility for PSG studies to be performed efficiently outside of the sleep laboratory at a larger scale, thus improving access for patients.
Competing Interest Statement
Conflict of interest. M. Stockhoff, M. Knoops-Borm and M. Tijssen are current employees of Onera Health. M. Knoops-Borm haves shareholding options in the company. Until recently, F. Raschella and M. Sekeri were employees of Onera Health. H. Schneider and S. Coughlin are contractors for Onera Health, receive compensation and have shareholdings in the company. H. Schneider receives compensation for studies conducted at the American Sleep Clinic. All other authors report no conflicts of interest. The study was managed by an independent CRO (Maxis Medical Inc, Frankfurt), who ensured regulatory compliance and compliance with ethical board requirements to prevent bias from COI.
Clinical Trial
NCT05310708
Clinical Protocols
https://clinicaltrials.gov/ct2/show/NCT05310708?term=NCT05310708&draw=2&rank=1
Funding Statement
Onera Health
Author Declarations
I confirm all relevant ethical guidelines have been followed, and any necessary IRB and/or ethics committee approvals have been obtained.
Yes
The details of the IRB/oversight body that provided approval or exemption for the research described are given below:
Ethics committee of Ethik-Kommission bei der Landesarztekammer Hessen gave ethical approval for this work.
I confirm that all necessary patient/participant consent has been obtained and the appropriate institutional forms have been archived, and that any patient/participant/sample identifiers included were not known to anyone (e.g., hospital staff, patients or participants themselves) outside the research group so cannot be used to identify individuals.
Yes
I understand that all clinical trials and any other prospective interventional studies must be registered with an ICMJE-approved registry, such as ClinicalTrials.gov. I confirm that any such study reported in the manuscript has been registered and the trial registration ID is provided (note: if posting a prospective study registered retrospectively, please provide a statement in the trial ID field explaining why the study was not registered in advance).
Yes
I have followed all appropriate research reporting guidelines and uploaded the relevant EQUATOR Network research reporting checklist(s) and other pertinent material as supplementary files, if applicable.
Yes
Data Availability
All data produced in the present work are contained in the manuscript.