Skip to main content
medRxiv
  • Home
  • About
  • Submit
  • ALERTS / RSS
Advanced Search

A comprehensive accuracy assessment of Samsung smartwatch heart rate and heart rate variability

View ORCID ProfileFatemeh Sarhaddi, Kianoosh Kazemi, Iman Azimi, Rui Cao, Hannakaisa Niela-Vilén, Anna Axelin, Pasi Liljeberg, Amir M. Rahmani
doi: https://doi.org/10.1101/2022.04.29.22274461
Fatemeh Sarhaddi
1Department of Computing, University of Turku, Turku, Finland
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
  • ORCID record for Fatemeh Sarhaddi
  • For correspondence: fasarh@utu.fi
Kianoosh Kazemi
1Department of Computing, University of Turku, Turku, Finland
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
Iman Azimi
1Department of Computing, University of Turku, Turku, Finland
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
Rui Cao
2Department of Electrical Engineering and Computer Science, University of California, Irvine, California, United States
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
Hannakaisa Niela-Vilén
3Department of Nursing Science, University of Turku, Turku, Finland
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
Anna Axelin
3Department of Nursing Science, University of Turku, Turku, Finland
4Department of Obstetrics and Gynaecology, Turku University Hospital and Faculty of Medicine, University of Turku, Turku, Finland
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
Pasi Liljeberg
1Department of Computing, University of Turku, Turku, Finland
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
Amir M. Rahmani
2Department of Electrical Engineering and Computer Science, University of California, Irvine, California, United States
5School of Nursing, University of California, Irvine, California, United States
6Institute for Future Health (IFH), University of California, Irvine, California, United States
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
  • Abstract
  • Full Text
  • Info/History
  • Metrics
  • Data/Code
  • Preview PDF
Loading

Abstract

Background Photoplethysmography (PPG) is a low-cost and easy-to-implement method to measure vital signs, including heart rate (HR) and heart rate variability (HRV). The method is widely used in various wearable devices. For example, Samsung smartwatches are PPG-based open-source wristbands used in remote well-being monitoring and fitness applications. However, PPG is highly susceptible to motion artifacts and environmental noise. A validation study is required to investigate the accuracy of PPG-based wearable devices in free-living conditions.

Objective We evaluate the accuracy of PPG signals – collected by the Samsung Gear Sport smartwatch in free-living conditions – in terms of HR and time-domain and frequency-domain HRV parameters against a medical-grade chest electrocardiogram (ECG) monitor.

Methods We conducted 24-hours monitoring using a Samsung Gear Sport smartwatch and a Shimmer3 ECG device. The monitoring included 28 participants (14 male and 14 female), where they engaged in their daily routines. We evaluated HR and HRV parameters during the sleep and awake time. The parameters extracted from the smartwatch were compared against the ECG reference. For the comparison, we employed the Pearson correlation coefficient, Bland-Altman plot, and linear regression methods.

Results We found a significantly high positive correlation between the smartwatch’s and Shimmer ECG’s HR, time-domain HRV, LF, and HF and a significant moderate positive correlation between the smartwatch’s and shimmer ECG’s LF/HF during sleep time. The mean biases of HR, time-domain HRV, and LF/HF were low, while the biases of LF and HF were moderate during sleep. The regression analysis showed low error variances of HR, AVNN, and pNN50, moderate error variances of SDNN, RMSSD, LF, and HF, and high error variances of LF/HF during sleep. During the awake time, there was a significantly high positive correlation of AVNN and a moderate positive correlation of HR, while the other parameters indicated significantly low positive correlations. RMSSD and SDNN showed low mean biases, and the other parameters had moderate mean biases. In addition, AVNN had moderate error variance while the other parameters indicated high error variances.

Conclusion The Samsung smartwatch provides acceptable HR, time-domain HRV, LF, and HF parameters during sleep time. In contrast, during the awake time, AVNN and HR show satisfactory accuracy, and the other HRV parameters have high errors.

Competing Interest Statement

The authors have declared no competing interest.

Funding Statement

This research was supported by the Academy of Finland https://www.aka.fi/en/ (Awards 316810 (AMR), and 316811 (AA)) and U.S. National Science Foundation https://www.nsf.gov/ (Awards CNS-1831918 and FW-HTF CNS-2026614) (AMR). The funders had no role in study design, data collection and analysis, decision to publish, or preparation of the manuscript

Author Declarations

I confirm all relevant ethical guidelines have been followed, and any necessary IRB and/or ethics committee approvals have been obtained.

Not Applicable

The details of the IRB/oversight body that provided approval or exemption for the research described are given below:

The study was conducted according to the ethical principles based on the Declaration of Helsinki and the Finnish Medical Research Act (No 488/1999). The study protocol received a favorable statement from the ethics committee (University of Turku, Ethics committee for Human Sciences, Statement no: 44/2019). The participants were informed about the study, both orally and in writing, before the written informed consent was obtained.

I confirm that all necessary patient/participant consent has been obtained and the appropriate institutional forms have been archived, and that any patient/participant/sample identifiers included were not known to anyone (e.g., hospital staff, patients or participants themselves) outside the research group so cannot be used to identify individuals.

Not Applicable

I understand that all clinical trials and any other prospective interventional studies must be registered with an ICMJE-approved registry, such as ClinicalTrials.gov. I confirm that any such study reported in the manuscript has been registered and the trial registration ID is provided (note: if posting a prospective study registered retrospectively, please provide a statement in the trial ID field explaining why the study was not registered in advance).

Not Applicable

I have followed all appropriate research reporting guidelines and uploaded the relevant EQUATOR Network research reporting checklist(s) and other pertinent material as supplementary files, if applicable.

Not Applicable

Footnotes

  • ↵* Fatemeh.Sarhaddi{at}utu.fi

Data Availability

The informed consent signed by the participants does not allow the data to be made publicly available due to ethical restriction. The data can be shared by signing a data use agreement. The data can be obtained from the corresponding author upon request.

Copyright 
The copyright holder for this preprint is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. It is made available under a CC-BY 4.0 International license.
Back to top
PreviousNext
Posted May 01, 2022.
Download PDF
Data/Code
Email

Thank you for your interest in spreading the word about medRxiv.

NOTE: Your email address is requested solely to identify you as the sender of this article.

Enter multiple addresses on separate lines or separate them with commas.
A comprehensive accuracy assessment of Samsung smartwatch heart rate and heart rate variability
(Your Name) has forwarded a page to you from medRxiv
(Your Name) thought you would like to see this page from the medRxiv website.
CAPTCHA
This question is for testing whether or not you are a human visitor and to prevent automated spam submissions.
Share
A comprehensive accuracy assessment of Samsung smartwatch heart rate and heart rate variability
Fatemeh Sarhaddi, Kianoosh Kazemi, Iman Azimi, Rui Cao, Hannakaisa Niela-Vilén, Anna Axelin, Pasi Liljeberg, Amir M. Rahmani
medRxiv 2022.04.29.22274461; doi: https://doi.org/10.1101/2022.04.29.22274461
Reddit logo Twitter logo Facebook logo LinkedIn logo Mendeley logo
Citation Tools
A comprehensive accuracy assessment of Samsung smartwatch heart rate and heart rate variability
Fatemeh Sarhaddi, Kianoosh Kazemi, Iman Azimi, Rui Cao, Hannakaisa Niela-Vilén, Anna Axelin, Pasi Liljeberg, Amir M. Rahmani
medRxiv 2022.04.29.22274461; doi: https://doi.org/10.1101/2022.04.29.22274461

Citation Manager Formats

  • BibTeX
  • Bookends
  • EasyBib
  • EndNote (tagged)
  • EndNote 8 (xml)
  • Medlars
  • Mendeley
  • Papers
  • RefWorks Tagged
  • Ref Manager
  • RIS
  • Zotero
  • Tweet Widget
  • Facebook Like
  • Google Plus One

Subject Area

  • Health Informatics
Subject Areas
All Articles
  • Addiction Medicine (230)
  • Allergy and Immunology (507)
  • Anesthesia (111)
  • Cardiovascular Medicine (1264)
  • Dentistry and Oral Medicine (207)
  • Dermatology (148)
  • Emergency Medicine (283)
  • Endocrinology (including Diabetes Mellitus and Metabolic Disease) (538)
  • Epidemiology (10056)
  • Forensic Medicine (5)
  • Gastroenterology (502)
  • Genetic and Genomic Medicine (2486)
  • Geriatric Medicine (240)
  • Health Economics (482)
  • Health Informatics (1653)
  • Health Policy (757)
  • Health Systems and Quality Improvement (638)
  • Hematology (250)
  • HIV/AIDS (538)
  • Infectious Diseases (except HIV/AIDS) (11896)
  • Intensive Care and Critical Care Medicine (627)
  • Medical Education (255)
  • Medical Ethics (75)
  • Nephrology (269)
  • Neurology (2304)
  • Nursing (140)
  • Nutrition (354)
  • Obstetrics and Gynecology (458)
  • Occupational and Environmental Health (537)
  • Oncology (1259)
  • Ophthalmology (377)
  • Orthopedics (134)
  • Otolaryngology (226)
  • Pain Medicine (158)
  • Palliative Medicine (50)
  • Pathology (326)
  • Pediatrics (737)
  • Pharmacology and Therapeutics (315)
  • Primary Care Research (282)
  • Psychiatry and Clinical Psychology (2295)
  • Public and Global Health (4850)
  • Radiology and Imaging (846)
  • Rehabilitation Medicine and Physical Therapy (493)
  • Respiratory Medicine (657)
  • Rheumatology (289)
  • Sexual and Reproductive Health (241)
  • Sports Medicine (228)
  • Surgery (273)
  • Toxicology (44)
  • Transplantation (131)
  • Urology (100)