Abstract
Objectives Numerous serologic immunoassays have been launched to detect antibodies to SARS-CoV-2, including rapid tests. Here, we validate use of a lateral flow immunoassay (LFI) intended for rapid screening and qualitative detection of anti-SARS-CoV-2 IgM and IgG in serum, plasma, and whole blood, and compare results with ELISA. We also seek to establish the value of LFI testing on blood obtained from a capillary blood sample.
Methods Samples collected by venous blood draw and capillary finger stick were obtained from patients with SARS-CoV-2 detected by RT-qPCR and control patients negative for SARS-CoV-2. Samples were tested with the 2019-nCoV IgG/IgM Detection Kit (Colloidal Gold) lateral flow immunoassay, and antibody calls were compared with results obtained by ELISA.
Results The Biolidics LFI kit shows clinical sensitivity of 92% at 7 days after PCR diagnosis of SARS-CoV-2 on venous blood. Test specificity was 92% for IgM and 100% for IgG. There was no significant difference in detecting IgM and IgG with Biolidics LFI and ELISA at D0 and D7 (p=1.00), except for detection of IgM at D7 (p=0.04). Finger stick whole blood of SARS-CoV-2 patients showed 93% sensitivity for antibody detection.
Conclusions Clinical performance of Biolidics 2019-nCoV IgG/IgM Detection Kit (Colloidal Gold) is comparable to ELISA and showed consistent results across different sample types. Furthermore, we show that capillary blood obtained by finger stick shows similar sensitivity for detecting anti-SARS-CoV-2 IgM and IgG antibodies as venous blood samples. This provides an opportunity for decentralized rapid testing in the community and may allow point-of-care and longitudinal self-testing for the presence of anti-SARS-CoV-2 antibodies.
Competing Interest Statement
The authors have declared no competing interest.
Funding Statement
Y. Feng's research is partially supported by NSF CAREER Grant DMS-2013789.
Author Declarations
All relevant ethical guidelines have been followed; any necessary IRB and/or ethics committee approvals have been obtained and details of the IRB/oversight body are included in the manuscript.
Yes
All necessary patient/participant consent has been obtained and the appropriate institutional forms have been archived.
Yes
I understand that all clinical trials and any other prospective interventional studies must be registered with an ICMJE-approved registry, such as ClinicalTrials.gov. I confirm that any such study reported in the manuscript has been registered and the trial registration ID is provided (note: if posting a prospective study registered retrospectively, please provide a statement in the trial ID field explaining why the study was not registered in advance).
Yes
I have followed all appropriate research reporting guidelines and uploaded the relevant EQUATOR Network research reporting checklist(s) and other pertinent material as supplementary files, if applicable.
Yes
Data Availability
There is no publicly available data associated with this study.