Abstract
Background Single-lead electrocardiograms (ECGs) can be recorded using widely available devices such as smartwatches and handheld ECG recorders. Such devices have been approved for atrial fibrillation (AF) detection. However, little evidence exists on the reliability of single-lead ECG interpretation. We aimed to assess the level of agreement on detection of AF by independent cardiologists interpreting single lead ECGs, and to identify factors influencing agreement.
Methods In a population-based AF screening study, adults aged ≥65 years old recorded four single-lead ECGs per day for 1-4 weeks using a handheld ECG recorder. ECGs showing signs of possible AF were identified by a nurse with the aid of an automated algorithm. These ECGs were reviewed by two independent cardiologists who assigned participant- and ECG-level diagnoses. Inter-rater reliability of AF diagnosis was calculated using linear weighted Cohen’s kappa (κw).
Results 185 participants and 1,843 ECGs were reviewed by both cardiologists. The level of agreement was moderate: κw = 0.42 (95% CI, 0.32 – 0.52) at the participant-level; and κw = 0.51 (0.46 – 0.56) at the ECG-level. At participant-level, agreement was associated with the number of adequate-quality ECGs recorded, with higher agreement in participants who recorded at least 67 adequate-quality ECGs. At ECG-level, agreement was associated with ECG quality and whether ECGs exhibited algorithm-identified possible AF.
Conclusions Inter-rater reliability of AF diagnosis from single-lead ECGs was found to be moderate in older adults. Strategies to improve reliability might include participant and cardiologist training and designing AF detection programmes to obtain sufficient ECGs for reliable diagnoses.
Clinical Trial Registration: ISRCTN 16939438; https://doi.org/10.1186/ISRCTN16939438
1. Introduction
The electrocardiogram (ECG) is a fundamental technique for assessing the functionality of the heart. The process for recording a 12-lead ECG was described 70 years ago (1), and to this day the 12-lead ECG remains widely used for the diagnosis and management of a range of heart conditions (2). Whilst the 12-lead ECG is highly informative, providing several ‘views’ of the heart’s electrical activity, it can only be measured by clinicians in a healthcare setting. Recently, clinical and consumer devices have become available which allow individuals to record a single-lead ECG on demand via a smartwatch or handheld device. This approach has a number of useful features: such ECGs can be measured by patients themselves with no clinical input, can be acquired synchronously with symptoms, can be repeated on multiple occasions with minimal inconvenience and can be transmitted electronically to healthcare providers (3).
Atrial fibrillation (AF) is a common arrhythmia which confers a fivefold increase in the risk of stroke (4) which can be mitigated through anticoagulation (5). A significant proportion of AF remains unrecognised (6) as it may be asymptomatic or occur only intermittently. Self-captured, single-lead ECGs could greatly assist in the detection of AF (7) when: (i) used by device owners, with ECGs acquired opportunistically, upon symptoms, or when prompted by a device (8); and when (ii) used in screening programmes, allowing multiple ECGs to be acquired from an individual over a period of weeks (9). Indeed, European Society of Cardiology guidelines support the use of single-lead ECGs acquired from wearable or mobile devices to identify AF (8). Whilst automated algorithms can be used to identify those ECGs which show evidence of AF and therefore warrant clinical review (10), a final diagnosis of AF must be made by a physician interpreting an ECG (8). To date, there is little evidence on the reliability of single-lead ECG interpretation for AF diagnosis, and most existing evidence is derived from ECGs collected from hospital patients (11–13).
We aimed to assess the level of agreement on detection of AF by independent cardiologists interpreting single lead ECGs, and to identify factors which influence agreement.
2. Methods
We assessed inter-rater agreement using ECG data collected in a population-based AF screening study, in which each participant recorded multiple ECGs. Agreement between cardiologist interpretations was assessed at the participant-level (i.e. the overall participant diagnosis) and the ECG-level (i.e. interpretations of individual ECGs). In addition, we investigated the influence of several factors on the level of agreement (e.g. participant age and ECG quality).
2.1 Data collection
We collected the data for these analyses in the SAFER (Screening for Atrial Fibrillation with ECG to Reduce stroke) Feasibility Study (ISRCTN 16939438), conducted in 2019 and approved by the London-Central Research Ethics Committee (REC ref: 18/LO/2066). Participants were older adults aged ≥ 65 years old, who were not receiving long-term anticoagulation for stroke prevention, not on the palliative care register, and not resident in a nursing home. All participants gave written informed consent.
In this study, older adults (aged 65 and over) recorded single-lead ECGs at home using the handheld Zenicor EKG-2 device (Zenicor Medical Systems AB) (10). This device measures a 30-second, single-lead ECG between the thumbs, using dry electrodes. Participants were asked to record four ECGs per day for either 1, 2 or 4 weeks. The ECGs were transferred to a central database for analysis and review.
Participant- and ECG-level diagnoses were obtained as follows (and as summarised in Figure 1). First, a computer algorithm was used to identify abnormal ECGs (Cardiolund ECG Parser algorithm, Cardiolund AB, Sweden). The algorithm has previously been found to have a sensitivity for AF detection of approximately 98% (10). Second, a nurse reviewed all the ECGs which were classified by the algorithm as abnormal, and manually corrected any algorithm misclassifications based on their clinical judgement. The nurse then identified participants for cardiologist review as those participants with at least one ECG classified as abnormal which the nurse deemed exhibited signs of possible AF (as detailed in (14)). Third, these participants were sent for review by two highly experienced cardiologists, both of whom had substantial ECG reviewing experience (GYHL and MRC). The cardiologists had access to all the ECGs from these participants, though it was not anticipated that ECGs that were classified as normal would be reviewed, or that all abnormal ECGs would be reviewed, once a participant-level diagnosis had been reached. Each cardiologist independently provided a diagnosis for each participant. For AF to be diagnosed it was required to be present for the whole 30 seconds, or the entire trace where the ECG was interpretable. No other formal definition of AF was provided for the cardiologists to use. In addition, on an ad hoc basis, the cardiologists also provided diagnoses for individual ECGs and labelled ECGs as ‘low-quality’. Diagnoses were categorised as: AF ≥30 seconds duration; cannot exclude AF; or, non-AF.
We extracted a subset of the collected data for the analysis as follows. Only data from those participants who were reviewed by both cardiologists were included in participant-level analyses. In addition, only those ECGs which were reviewed by both cardiologists were included in ECG-level analyses. We excluded from analyses any ECGs for which a cardiologist’s initial diagnosis was not recorded (prior to subsequent resolution of disagreements).
2.2 Data Processing
We obtained the characteristics of each ECG as follows: First, the computer algorithm extracted the following characteristics: heart rate, ECG quality (either normal or poor quality), level of RR-interval variability (calculated as the standard deviation of RR-intervals divided by the mean RR-interval), and whether or not an ECG exhibited algorithm-identified possible AF (defined as the ECG having either irregular RR-intervals or a fast regular heart rate). Second, the quality of ECGs was obtained by combining the quality assessment provided by the algorithm with cardiologist comments on ECG quality: any ECGs which the algorithm or at least one cardiologist deemed to be of poor quality were classed as low quality in the analysis. ECGs for which the algorithm was unable to calculate heart rate or RR-interval variability were excluded from analyses requiring those characteristics.
2.3 Statistical Analysis
We assessed the reliability of ECG interpretation using both participant-level diagnoses and ECG-level cardiologist diagnoses. First, we reported the overall levels of agreement. Second, we assessed the influence of different factors on levels of agreement, such as the influence of ECG quality. The factors assessed at the participant-level were: age, gender, number of adequate-quality ECGs recorded by a participant, and the number of ECGs recorded by a participant exhibiting algorithm-identified possible AF. The factors assessed at the ECG-level were: heart rate, RR-interval variability, ECG quality, and whether or not an ECG exhibited algorithm-identified possible AF. We investigated factors which were continuous variables (such as heart rate) by grouping values into categories with similar sample sizes (e.g. heart rates were categorised as 30-59 bpm, 60-69 bpm, etc).
We assessed agreement between cardiologists using inter-rater reliability statistics. The primary statistic, Cohen’s kappa, κ, provides a measure of the difference between the actual level of agreement between cardiologists, and the level of agreement that would be expected by random chance alone. Values for κ range from −1 to 1, with −1 indicating complete disagreement; 0 the level expected by chance; 0.01-0.20 slight agreement; 0.21-0.40 fair agreement; 0.41-0.60 moderate agreement; 0.61-0.80 substantial agreement; 0.81-0.99 almost perfect agreement; and 1 perfect agreement (15). The second statistic, a weighted Cohen’s kappa, κw, reflects the greater consequences of a disagreement of ‘AF’ vs ‘non-AF’, compared to a disagreement of ‘cannot exclude AF’ vs either ‘AF’ or ‘non-AF’. We weighted disagreements of ‘AF’ vs ‘non-AF’ as complete disagreements, whereas disagreements including ‘cannot exclude AF’ were weighted equivalently to the level expected by chance. We reported the third statistic, percentage agreement, to facilitate comparisons with previous studies.
We calculated 95% confidence intervals for κ and κw using bootstrapping. We undertook tests for significant associations between factors (e.g. heart rate) and the level of agreement using a chi-square test for independence between the proportion of agreement in each category.
3. Results
Of the 2,141 participants who were screened, 190 had ECGs which underwent cardiologist review and were therefore included in the participant-level analyses, as shown in Figure 1. Most participants’ ECGs were not sent for cardiologist review (1,951 participants) because either: (i) the computer algorithm did not find any abnormalities in their ECGs (603 participants); or (ii) the nurse reviewer judged that none of their abnormal ECGs exhibited signs of possible AF (1,348 participants).
The 190 participants whose ECGs underwent cardiologist review recorded a total of 15,258 ECGs, with a median (lower – upper quartiles) of 67.0 (56.0 - 112.0) ECGs each. The two cardiologists assigned diagnoses to 1,996 and 4,411 of these ECGs respectively of which 1,872 ECGs were assigned diagnoses by both cardiologists. Initial diagnoses (prior to subsequent resolution of disagreements) were not recorded for 29 of these ECGs, leaving 1,843 available for ECG-level analyses (see Figure 2).
3.1. Reliability of AF diagnosis at the participant-level
The inter-rater reliability of AF diagnosis at the participant-level, when the cardiologists had access to all the ECGs recorded by a participant, was moderate (κw = 0.48 (0.37 − 0.58); κ = 0.42 (0.32 − 0.52); and %agree= 66.3% ( Table 1).
The results for the relationship between the level of agreement between cardiologists and factors at the participant- and ECG-level are presented in Figure 3 and Table 2. At the participant-level, the level of agreement was significantly associated with the number of adequate-quality ECGs recorded by a participant. Participants who recorded at least 67 adequate-quality ECGs had a significantly higher level of agreement in their diagnoses than those who recorded fewer than 67. There was agreement on 52.6% of participant-level diagnoses in those participants with <67 adequate-quality ECGs, compared to 80.0% in those with 67 or more. Of the 31 participants for whom there was complete disagreement (where one cardiologist diagnosed AF and the other diagnosed non-AF), 23 (74%) recorded <67 adequate-quality ECGs. There was no significant association between the level of agreement and age or gender.
3.2. Reliability of ECG interpretation
The inter-rater reliability of AF diagnosis at the individual ECG-level was moderate (κw = 0.58 (0.53 − 0.63); κ = 0.51 (0.46 − 0.56); and %agree = 86.1% (Table 3). Referring to the ECG-level results in Figure 3 and Table 2, the level of agreement was significantly associated with ECG quality, with low-quality ECGs associated with a lower level of agreement. This remained regardless of whether quality was assessed using cardiologist comments on ECG quality, the automated algorithm assessment, or a combination of both. The level of agreement was also significantly associated with whether or not an ECG exhibited algorithm-identified possible AF, where ECGs exhibiting possible AF were associated with a higher level of agreement. The was no significant association between the level of agreement and heart rate or RR-interval variability.
3.3. Comparison of cardiologists’ reviewing practices
The two cardiologists’ reviewing practices differed. At the participant-level, one cardiologist diagnosed more participants with AF than the other (72 out of 190, i.e. 38%, vs. 50, i.e. 26%) (see Table 1). Similarly, at the ECG-level, this cardiologist diagnosed more ECGs as AF than the other (235 out of 1,843, i.e. 13%, vs. 179, i.e. 9.7%), and more ECGs as ‘cannot exclude AF’ than the other (119, i.e. 6%, vs. 63, i.e. 3%) (see Table 3). Most of the ECGs diagnosed as AF by the cardiologists exhibited an irregular rhythm as identified by the algorithm (95% of the 235 ECGs diagnosed as AF by one cardiologist, 88% of the 179 ECGs diagnosed as AF by the other cardiologist, and 95% of the 137 ECGs diagnosed as AF by both cardiologists).
4. Discussion
4.1. Summary of findings
This study provides evidence on the inter-rater reliability of single-lead electrocardiogram interpretation, and the factors that influence this. Moderate agreement was observed between cardiologists on participant-level diagnoses of AF in a population-based AF screening study when this diagnosis was made using multiple ECGs per participant. The key factor associated with the level of agreement at the participant-level was the number of adequate-quality ECGs recorded by a participant, with higher levels of agreement in those who recorded more adequate-quality ECGs. Moderate agreement was observed between cardiologists on the diagnoses of individual ECGs. Similarly, at the ECG-level, low-quality ECGs were associated with lower levels of agreement. In addition, lower levels of agreement were observed on those ECGs not exhibiting algorithm-identified possible AF.
4.2. Comparison with existing literature
The levels of agreement in AF diagnosis from single-lead ECGs observed in this study are lower than in many previous studies. Previous studies have found almost perfect agreement when interpreting 12-lead ECGs, but lower levels of agreement when interpreting single-lead ECGs. In an analysis of 12-lead ECGs from the SAFE AF Screening Trial, cardiologists agreed on the diagnosis of 99.7% of ECGs (all but 7 of 2,592 analysed ECGs)(16). In comparison, in the present study of single-lead ECGs cardiologists agreed on the diagnosis of 86.1% of ECGs (1,587 out of 1,843 ECGs). However, the proportion of normal ECGs included in this study was substantially lower than in the SAFE AF Screening Trial (less than 1% in this study, versus 93% in SAFE), so the simple level of agreement is not directly comparable. Similarly, in a study of the diagnosis of supraventricular tachycardia in hospital patients, an almost perfect agreement of κ = 0.97 was observed in interpretation of 12-lead ECGs, compared to a substantial agreement of κ = 0.76 when using single-lead ECGs from the same patients (17). The previously reported levels of agreement for the diagnosis of AF from single-lead ECGs have varied greatly between studies: fair agreement was observed by Kearley et al. (18) (κ = 0.28); moderate agreement was observed by Lowres et al. (19) (weighted κ = 0.4); substantial agreements were observed by Poulsen et al. (12) (κ = 0.65) and Kearley et al. (18) (κ = 0.76); and almost perfect agreements were observed by Desteghe et al. (11)(κ = 0.69 to 0.86), Koshy et al. (20) (κ = 0.80 to 0.83), Wegner et al. (13) (κ = 0.90), and Racine et al. (21)(κ = 0.94). The variation in levels of agreement may have been contributed to by study setting and underlying frequency of AF, since those studies which reported the lowest levels of agreement took place out-of-hospital (18, 19). The present study, conducted in the community, similarly observed lower levels of agreement than many other studies (κ = 0.42 at the participant-level, and κ = 0.51 at the ECG-level). In the context of AF screening, a 69.2% level of agreement has been reported in a previous AF screening study by Pipilas et al. (22), compared to 86.1% in the present screening study. The low levels of agreement in the present study could have been contributed to by: (i) the ECGs being more challenging to review as an algorithm and a nurse filtered out most ECGs which did not exhibit signs of AF (and are therefore easier to interpret) prior to cardiologist review; (ii) the ECGs being of lower quality since participants recorded ECGs themselves without clinical supervision; and (iii) the use of an additional diagnostic category of ‘cannot exclude AF’.
This study’s findings about factors which influence the reliability of ECG interpretation complement those reported previously (11,12,13,22,23). It has previously been reported that ECGs exhibiting baseline wander, noise, premature beats, or low-amplitude atrial activity are associated with mis-diagnoses (11,23). In this study low-quality ECGs were similarly associated with lower levels of agreement between cardiologists. The significant proportion of low-quality ECGs obtained when using a handheld ECG device has been reported previously, with 12% of ECGs being judged as ‘very low quality’ in (22), 13% as ‘not useable’ in (12), and 20% as ‘inadequate quality’ in (13).
The accuracy of both automated and manual diagnosis of AF from single-lead ECGs has been assessed previously. A recent meta-analysis found pooled sensitivities and specificities of automated ECG diagnoses of 89% and 99% respectively in the community setting (24). The accuracy of manual diagnoses has varied greatly between previous studies, with sensitivities and specificities in comparison to reference 12-lead ECGs reported as: 77.4% and 73.0% (22), 90% and 79% (21), 76-92% and 84-100% (20), 89-100% and 85-88% (25), 92.5% and 89.8% (26), 93.9% and 90.1% (18), 100% and 94% (13), and 100% and 100% (27). In all of these studies, the single-lead ECGs were recorded under supervision. In contrast, the present study considered ECGs collected using a telehealth device at home without supervision.
4.3. Strengths and limitations
There are several strengths to this study. First, we assessed the level of agreement in both participant-level ECG-level AF diagnoses which is of particular relevance in AF screening, whereas most previous work has been limited to ECG-level diagnoses. Second, the ECGs used in this study were collected in a prospective population-based AF screening study, and are therefore representative of ECGs captured in telehealth settings by older adults without clinical supervision. The ECGs were recorded using dry electrodes, as opposed to the gel electrodes used in clinical settings. Dry electrodes can result in poorer conduction and therefore lower signal quality, making interpretation more challenging. Since smartwatches also use dry electrodes, the findings are expected to be relevant to the growing use of ECG-enabled consumer devices. Third, the ECGs included in the analysis are representative of those which would be sent for clinical review in real-world settings: ECGs without signs of abnormalities were excluded using an automated, CE-marked analysis system, leaving only those ECGs with signs of abnormalities for review. Fourth, the study included a large number of ECGs (1,843), each interpreted by two cardiologists. Fifth, we used Cohen’s kappa statistic to assess the level of agreement between cardiologists: this statistic takes into account agreement by chance unlike the percentage agreement (Viera and Garrett, 2005).
The key limitations to this study are as follows. First, the findings are based on data from only 190 participants who had an abnormal ECG flagged in the study. Second, inter-rater agreement was assessed using diagnoses provided by only two cardiologists. Third, not all ECGs sent for review were interpreted by both cardiologists, with those not interpreted by both cardiologists excluded from the analysis. Fourth, the initial diagnosis was not recorded for a small minority of the ECGs reviewed by both cardiologists (29 out of 1,872, 1.5%), so these were not included in the analysis. Finally, it should be remembered that the study assessed the reliability of ECG interpretation (i.e. the level of agreement between two cardiologists), rather than the accuracy of ECG interpretation (i.e. a comparison of cardiologist interpretation against an independent reference). In doing so, the study identified factors associated with reduced levels of agreement, providing evidence on how to improve the level of agreement, and subsequently the reliability of interpretation.
4.4. Implications
This study indicates that steps should be taken to ensure diagnoses based on single-lead ECGs are as reliable as possible. Out of 2,141 participants screened for AF, there was agreement between cardiologists on diagnoses of AF for 44 participants, complete disagreement for 31 participants (AF vs. non-AF), and partial disagreement for 33 participants (AF or non-AF vs. cannot exclude AF). In terms of disease prevalence, there was agreement on AF diagnosis in 2.1% of the sample population, complete disagreement in 1.4%, and partial disagreement in 1.5%.
The findings could inform the design of AF screening programmes. AF screening programmes often include collection of multiple short ECGs (or a continuous ECG recording) over a prolonged period to capture even infrequent episodes of paroxysmal AF. The results of this study indicate that a prolonged period is also required to obtain reliable diagnoses: at least 67 adequate-quality ECGs were required for a reliable diagnosis in this study, providing evidence that screening programmes should be designed to capture at least this many adequate-quality ECGs from all participants (i.e. at least 17 days of screening when recording 4 ECGs per day, and potentially 21 days of screening to account for missed or low-quality ECGs). In addition, no association was found between participant gender or age and the reliability of diagnoses, indicating that it is reasonable to use single-lead ECGs in older adults of a wide range of ages (from 65 to 90+ in this study).
The findings of this study could also underpin strategies to obtain more reliable participant-level diagnoses through personalised screening. Those individuals who are likely to receive a less reliable diagnosis could be identified by using an automated algorithm to analyse the quality of incoming ECGs, and then the duration of screening could be extended in those individuals without sufficient adequate-quality ECGs. This could help increase reliability by increasing the number of adequate-quality ECGs available for diagnosis. Second, participants with a high proportion of low-quality ECGs could be offered additional training on ECG measurement technique, potentially by telephone.
This study highlights the need to ensure single-lead ECG interpreters receive sufficient training. The ECGs in this study were interpreted by highly experienced cardiologists, and yet there was still disagreement over diagnoses for 16% of those participants sent for cardiologist review. If single-lead ECG-based AF screening is widely adopted in the future, then it will be important to ensure all ECG interpreters receive sufficient training and gain sufficient experience in single-lead ECG interpretation to provide reliable diagnoses. We note that single-lead ECG interpretation presents additional challenges beyond those encountered in 12-lead ECG interpretation: ECGs may be of lower quality (12), P-waves may not be as visible (11), and only one lead is available.
The findings of this study indicate that it is important that the quality of single-lead ECGs is as high as possible. Particularly given the implications of an AF diagnosis such as recommendations for anticoagulation treatment which increases the risk of bleeding. The development of consumer and telehealth ECG devices involves making a range of design decisions which can influence the quality of ECGs sent for clinical review, including: the size, type, and anatomical position of electrodes; the filtering applied to signals to reduce noise; and whether to exclude ECGs of insufficient quality from clinical review (and if so, how best to identify these ECGs). Device should designers consider the potential effects of these design decisions on the reliability of diagnoses.
5. Conclusion
Moderate agreement was found between cardiologists when diagnosing AF from single-lead ECGs in an AF screening study. The study indicates that for every 100 screening participants diagnosed with AF by two cardiologists, there would be complete disagreement over the diagnosis of 70 further participants. This provides great incentive for ensuring that the interpretation of single-lead ECGs is as reliable as possible. Key factors were identified which influence the reliability of single-lead ECG interpretation. Most importantly, the quality of ECG signals greatly influenced reliability. In addition, when multiple ECGs were acquired from an individual, the reliability of participant-level diagnoses was influenced by the number of adequate-quality ECGs available for interpretation. This new evidence could help improve single-lead ECG interpretation, and consequently increase the effectiveness of screening for AF using single-lead ECG devices. Future work should investigate how to obtain ECGs of the highest possible quality in the telehealth setting, and how best to train ECG interpreters to ensure diagnoses are as accurate as possible.
Data Availability
Requests for pseudonymised data should be directed to the SAFER study co-ordinator (Andrew Dymond using SAFER{at}medschl.cam.ac.uk) and will be considered by the investigators, in accordance with participant consent.
Acknowledgment
This study is funded by the National Institute for Health and Care Research (NIHR), Programme Grants for Applied Research Programme (Reference Number RP-PG0217-20007); the NIHR School for Primary Care Research (SPCR-2014-10043, project 410); and the British Heart Foundation (BHF) grant number FS/20/20/34626. The views expressed are those of the authors and not necessarily those of the NIHR or the Department of Health and Social Care.
Footnotes
Funding: This study is funded by the National Institute for Health Care Research (NIHR), grant number RP-PG-0217-20007; and the British Heart Foundation (BHF), grant number FS/20/20/34626. The views expressed are those of the authors and not necessarily those of the NIHR or the Department of Health and Social Care.
Disclosures: MRC is employed by Astrazeneca PLC. BF has received speaker fees, honoraria, and non-financial support from the BMS and Pfizer Alliance; and loan devices for investigator initiated studies from Alivecor: all were unrelated to the present study, but related to screening for AF. SJG has received honoraria from Astra Zeneca and Eli Lilly for contributing to postgraduate education concerning type 2 diabetes to specialists and primary care teams. FDRH reports occasional consultancy for BMS/Pfizer, Bayer and BI over the last five years. HCL is employed by Zenicor Medical Systems AB. GYHL: Consultant and speaker for BMS/Pfizer, Boehringer Ingelheim, Daiichi-Sankyo, Anthos. No fees are received personally. JM has performed consultancy work for BMS/Pfizer and Omron. PHC has performed consultancy work for Cambridge University Technical Services, and has received honoraria from IOP Publishing and Emory University (the latter not received personally).
Clarification of the meaning of 'possible AF'.