Dual- and Multi-Energy CT: Principles, Technical Approaches, and Clinical Applications

Radiology. 2015 Sep;276(3):637-53. doi: 10.1148/radiol.2015142631.

Abstract

In x-ray computed tomography (CT), materials having different elemental compositions can be represented by identical pixel values on a CT image (ie, CT numbers), depending on the mass density of the material. Thus, the differentiation and classification of different tissue types and contrast agents can be extremely challenging. In dual-energy CT, an additional attenuation measurement is obtained with a second x-ray spectrum (ie, a second "energy"), allowing the differentiation of multiple materials. Alternatively, this allows quantification of the mass density of two or three materials in a mixture with known elemental composition. Recent advances in the use of energy-resolving, photon-counting detectors for CT imaging suggest the ability to acquire data in multiple energy bins, which is expected to further improve the signal-to-noise ratio for material-specific imaging. In this review, the underlying motivation and physical principles of dual- or multi-energy CT are reviewed and each of the current technical approaches is described. In addition, current and evolving clinical applications are introduced.

Publication types

  • Research Support, N.I.H., Extramural
  • Review

MeSH terms

  • Algorithms
  • Humans
  • Radiography, Dual-Energy Scanned Projection*
  • Tomography, X-Ray Computed / methods*