Whole-brain susceptibility mapping at high field: a comparison of multiple- and single-orientation methods

Neuroimage. 2010 Nov 1;53(2):515-25. doi: 10.1016/j.neuroimage.2010.06.070. Epub 2010 Jul 6.

Abstract

Optimisation and comparison of the performance of three different methods for calculating three-dimensional susceptibility maps of the whole brain from gradient-echo (phase and modulus) image data acquired at 7 T is described. The methods studied are a multiple-orientation method in which image data acquired with the head at several different angles to the main field are combined and two methods which use data acquired at a single orientation: the first of these is based on exclusion of some k-space data from the calculation (through thresholding of the dipolar field kernel), while the second incorporates a regularisation method that is based on using information from the modulus images. The methods were initially optimised via analysis of data from a phantom containing different compartments of known susceptibility. As part of this work, a novel high-pass filtering methodology was introduced to remove background fields from field maps based on phase data. The optimised methods were successfully applied to high-resolution (0.7 mm isotropic) whole-brain modulus and phase data acquired in vivo from five healthy male subjects, 25-30 years of age. The multiple-orientation method yielded high quality susceptibility maps, out-performing the single-orientation methods. Venous blood vessels as well as the substantia nigra and globus pallidus brain regions showed particularly high positive susceptibility offsets relative to surrounding tissue, consistent with high deoxyhemoglobin and non-heme iron content, respectively. To compare the performance of the different methods, regions of interest were drawn in deep grey matter structures and in cortical grey and white matter. The threshold-based approach was fast and simple to use, but underestimated susceptibility differences and showed significant artefacts due to noise amplification in difficult regions of k-space. The regularised single-orientation method yielded contrast dependent on the choice of spatial priors, but demonstrated the potential to yield susceptibility maps of a similar quality to those calculated using data acquired at multiple orientations to the field.

Publication types

  • Comparative Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adult
  • Algorithms
  • Brain / physiology*
  • Brain Chemistry / physiology
  • Brain Mapping / methods*
  • Cerebrovascular Circulation
  • Globus Pallidus / anatomy & histology
  • Globus Pallidus / blood supply
  • Hemoglobins / metabolism
  • Humans
  • Image Processing, Computer-Assisted / methods*
  • Magnetic Resonance Imaging / methods*
  • Male
  • Nonheme Iron Proteins / metabolism
  • Phantoms, Imaging
  • Reproducibility of Results
  • Substantia Nigra / anatomy & histology
  • Substantia Nigra / blood supply

Substances

  • Hemoglobins
  • Nonheme Iron Proteins
  • deoxyhemoglobin