Gastroenterology

Gastroenterology

Volume 134, Issue 4, April 2008, Pages 960-974.e8
Gastroenterology

Clinical–Liver, Pancreas, and Biliary Tract
Performance of Transient Elastography for the Staging of Liver Fibrosis: A Meta-Analysis

https://doi.org/10.1053/j.gastro.2008.01.034Get rights and content

Background & Aims: Transient elastography has been studied in a multitude of liver diseases for the staging of liver fibrosis with variable results. A meta-analysis was performed to assess the overall performance of transient elastography for the diagnosis of liver fibrosis and to analyze factors influencing the diagnostic accuracy. Methods: Literature databases and international conference abstracts were searched. Inclusion criteria were as follows: evaluation of transient elastography, liver biopsy as reference, and assessment of the area under the receiver operating characteristic curve (AUROC). The meta-analysis was performed using the random-effects model for the AUROC, summary receiver operating curve techniques, as well as meta-regression approaches. Results: Fifty studies were included in the analysis. The mean AUROC for the diagnosis of significant fibrosis, severe fibrosis, and cirrhosis were 0.84 (95% confidence interval [CI], 0.82–0.86), 0.89 (95% CI, 0.88–0.91), and 0.94 (95% CI, 0.93–0.95), respectively. For the diagnosis of significant fibrosis a significant reduction of heterogeneity of the AUROC was found when differentiating between the underlying liver diseases (P < .001). Other factors influencing the AUROC were the scoring system used and the country in which the study was performed. Age, body mass index, and biopsy quality did not have a significant effect on the AUROC. Conclusions: Transient elastography can be performed with excellent diagnostic accuracy and independent of the underlying liver disease for the diagnosis of cirrhosis. However, for the diagnosis of significant fibrosis, a high variation of the AUROC was found that is dependent on the underlying liver disease.

Section snippets

Transient Elastography

Transient elastography is a novel method. The first clinical data from transient elastography were published in 2002. Transient elastography is performed with an ultrasound transducer probe mounted on the axis of a vibrator. A vibration transmitted from the vibrator toward the tissue induces an elastic shear wave that propagates through the tissue. These propagations are followed by pulse-echo ultrasound acquisitions and their velocity is measured, which is related directly to tissue stiffness.

Results

The literature search yielded 56 full-length articles and 123 abstracts that evaluated transient elastography. They were read in full. Fifty studies were included in the meta-analysis according to the inclusion and exclusion criteria. In detail, these were 15 full-length articles6, 7, 16, 17, 18, 19, 20, 21, 22, 23, 24, 25, 26, 27, 28 and 35 abstracts.29, 30, 31, 32, 33, 34, 35, 36, 37, 38, 39, 40, 41, 42, 43, 44, 45, 46, 47, 48, 49, 50, 51, 52, 53, 54, 55, 56, 57, 58, 59, 60, 61, 62, 63

The

Discussion

The systematic literature search revealed 50 studies evaluating the diagnostic performance of transient elastography for the staging of liver fibrosis, which fulfilled the inclusion criteria and reported enough data to perform a meta-analysis. The aim of the systematic literature search was to include all relevant publications (including abstracts) with the main focus on the meta-analysis of the AUROC. A meta-analysis based on individual data was not the scope of the present study. Therefore,

References (74)

  • M. Fraquelli et al.

    Reproducibility of transient elastography in the evaluation of liver fibrosis in patients with chronic liver disease

    Gut

    (2007)
  • J.J. Deeks

    Systematic reviews of evaluation of diagnostic and screening tests

    BMJ

    (2001)
  • R.B. Haynes et al.

    Optimal search strategies for retrieving scientifically strong studies of diagnosis from Medline: analytical survey

    BMJ

    (2004)
  • T. Poynard et al.

    Variability of the area under the receiver operating characteristic curves in the diagnostic evaluation of liver fibrosis markers: impact of biopsy length and fragmentation

    Aliment Pharmacol Ther

    (2007)
  • T. Poynard et al.

    Standardization of ROC curve areas for diagnostic evaluation of liver fibrosis markers based on prevalences of fibrosis stages

    Clin Chem

    (2007)
  • P. Whiting et al.

    The development of QUADAS: a tool for the quality assessment of studies of diagnostic accuracy included in systematic reviews

    BMC Med Res Methodol

    (2003)
  • B. Littenberg et al.

    Estimating diagnostic accuracy from multiple conflicting reports: a new meta-analytic method

    Med Decis Making

    (1993)
  • J.A. Carrion et al.

    Transient elastography for diagnosis of advanced fibrosis and portal hypertension in patients with hepatitis C recurrence after liver transplantation

    Liver Transpl

    (2006)
  • C. Colletta et al.

    Value of two noninvasive methods to detect progression of fibrosis among HCV carriers with normal aminotransferases

    Hepatology

    (2005)
  • C. Corpechot et al.

    Assessment of biliary fibrosis by transient elastography in patients with PBC and PSC

    Hepatology

    (2006)
  • V. De Ledinghen et al.

    Diagnosis of hepatic fibrosis and cirrhosis by transient elastography in HIV/hepatitis C virus-coinfected patients

    J Acquir Immune Defic Syndr

    (2006)
  • A. Erhardt et al.

    Transient elastography for diagnosing liver cirrhosis

    Dtsch Med Wochenschr

    (2006)
  • N. Ganne-Carrie et al.

    Accuracy of liver stiffness measurement for the diagnosis of cirrhosis in patients with chronic liver diseases

    Hepatology

    (2006)
  • E. Gomez-Dominguez et al.

    Transient elastography: a valid alternative to biopsy in patients with chronic liver disease

    Aliment Pharmacol Ther

    (2006)
  • P. Nahon et al.

    Liver stiffness measurement versus clinicians’ prediction or both for the assessment of liver fibrosis in patients with chronic hepatitis C

    Am J Gastroenterol

    (2006)
  • M. Ziol et al.

    Noninvasive assessment of liver fibrosis by measurement of stiffness in patients with chronic hepatitis C

    Hepatology

    (2005)
  • J. Foucher et al.

    Diagnosis of cirrhosis by transient elastography (FibroScan): a prospective study

    Gut

    (2006)
  • T. Takeda et al.

    Usefulness of noninvasive transient elastography for assessment of liver fibrosis stage in chronic hepatitis C

    World J Gastroenterol

    (2006)
  • C. Baldaia et al.

    Transient elastography in chronic hepatitis C—comparison between different noninvasive methods for liver fibrosis assessment

    Hepatology

    (2006)
  • C. Barrault et al.

    Non invasive assessment of liver graft fibrosis by transient elastography after liver transplantation

    Hepatology

    (2005)
  • P.L. Blanc et al.

    Liver stiffness in chronic hepatitis C: will it modify the assessment of patients?

    J Hepatol

    (2007)
  • L. Castera et al.

    Independent validation and comparison with FibroScan, FibroTest and liver biopsy of clinical glycomics for the non invasive assessment of liver fibrosis in chronic hepatitis C

    Hepatology

    (2005)
  • L. Castera et al.

    Prospective and independent validation of the Lok index for prediction of cirrhosis in patients with chronic hepatitis C: comparison with FibroScan, FibroTest and APRI

    DDW

    (2006)
  • Chang J, Tan HH, Yew BS, et al. Transient elastography (FibroScan®) to assess hepatic fibrosis in Chinese with chronic...
  • E. Chanteloup et al.

    FibroScan® is a new non invasive method for the detection of cirrhosis and its complications in patients with chronic liver diseases. Results of a prospective study in 456 patients

    Hepatology

    (2004)
  • B. Coco et al.

    Monitoring liver stiffness: a new tool to measure liver fibrosis during therapy

    Hepatology

    (2005)
  • B. Coco et al.

    Liver stiffness measured by transient elastography: the influence of biochemical activity

    J Hepatol

    (2006)
  • Cited by (1271)

    • Differentiation of bladder cancer with water flow elastography (WaFE)

      2024, Journal of the Mechanical Behavior of Biomedical Materials
    View all citing articles on Scopus

    This study was supported by the Federal Ministry of Education and Research (BMBF) program Kompetenznetz Hepatitis (Hep-Net).