The Journal of Steroid Biochemistry and Molecular Biology
Fast and sensitive liquid chromatography–mass spectrometry assay for seven androgenic and progestagenic steroids in human serum
Highlights
► LC–MS/MS method for seven human serum steroids was developed. ► Oxime derivatization used for increased method sensitivity. ► Total run time 7 min with quantification limits between 0.03 and 0.34 nM. ► Method was validated and used for serum steroid profiling of endometriosis patients.
Introduction
Steroid hormones are endogenous compounds that are derived from cholesterol by a series of enzymatic reactions in the adrenal glands, gonads, and placenta. The steroids typically possess strong physiological effects at very low concentrations via binding to nuclear receptors at nano- or picomolar levels, and variations in the concentrations, e.g. of estradiol, progesterone, and testosterone are routinely used for diagnostic purposes. Although assaying a single steroid in serum or plasma can be a useful diagnostic parameter, in many cases the relative concentrations of several steroids and their metabolites are critical [1]. Thus, the simultaneous quantitative measurement of circulating steroids is expected to be an important part of the research, diagnosis, and treatment of the disorders of steroid hormone synthesis and metabolism in the future.
Classical methods for plasma and serum steroid analyses are based on immunoassays (IA) and related techniques. Although still in use, it is now known that many IAs suffer from specificity issues and often overestimate the concentrations, particularly at the lower end of the concentration range [2], [3], [4], [5]. Also, the interassay precision of many IAs is poor [6] and multiple assays are needed for steroid profiling. For these reasons, the use of mass spectrometry-based methods for clinical steroid analysis is rapidly increasing [7], [8].
While mass spectrometry has a long history in steroid analysis in the form of GC–MS, the current trend is towards LC–MS because of its potential for higher throughput and simpler sample preparation [8], [9]. However, although steroids are well suited for liquid chromatography, their structural properties potentially lead to poor ionization efficiency, compromising the assay sensitivity. The weak gas-phase proton affinity of most steroids does not favor the use of positive-mode ESI, as different adducts are usually generated instead of [M+H]+ ions [10], [11]. The three major estrogens (estrone, estradiol, and estriol) have at least one phenolic hydroxyl group, which can be ionized with negative ESI [12], but the vast majority of steroids lack this property. To increase the assay sensitivity for these compounds, APCI or APPI have been used instead of ESI [13], [14], [15]. However, regardless of the ionization technique, improvements in the assay sensitivity have been realized after chemical derivatization of the analytes [10], [11], [16], [17], [18], [19].
Although derivatization is shown to improve the ionization of many steroids, it can also complicate the sample preparation process. This can reduce the reliability of the method, usually due to the uncontrollable reaction recovery and the additional purification procedures needed for the reaction mixture [1]. However, some derivatization techniques, such as the recently introduced oxime derivatization of ketosteroids [10], [20], [21], enable the injection of the reaction mixture into the separation column as such [3], and the inclusion of stable isotope-labeled internal standards will effectively compensate any variability in the reaction recovery. The derivatization technique also significantly increases the ESI response of neutral keto- or oxosteroids [10], [20].
The aim of the presented work was to develop a method for fast and reliable quantification of the major androgens and progestagens in the biosynthetic pathway from cholesterol to testosterone. Although various LC–MS assays for these compounds in serum have been published [22], [23], [24], [25], [26], [27], [28], [29], [30], [31], methods for their simultaneous determination, with sufficient sensitivity for the analysis of both female and male samples have not been previously described. The presented assay employs oxime derivatization and positive-mode ESI, with a simplified sample preparation technique that incorporates the derivatization of the analytes into the sample extraction process.
Section snippets
Materials, reagents, and equipment
Standard compounds 17α-hydroxypregnenolone, 17α-hydroxyprogesterone, androstenedione, dehydroepiandrosterone (DHEA), testosterone, and pregnenolone were obtained from Sigma–Aldrich (St. Louis, MO). Progesterone was from Steraloids (Newport, RI). Internal standards d7-androstenedione and d9-progesterone were from Steraloids, d4-pregnenolone, d8-17α-hydroxyprogesterone, and d3-17α-hydroxypregnenolone from C/D/N Isotopes (Quebec, Canada), d3-testosterone and d6-dehydroepiandrosterone from
Chromatography
During oxime derivatization of ketosteroids, two geometric stereoisomers may be formed, with the hydroxyl group in either syn- or anti-position. These isomers can be separated chromatographically [36], which is usually undesirable as it decreases the overall response achievable from a single derivatized compound. On the other hand, adequate chromatographic separation of structurally similar, isobaric steroids must be achieved. For these reasons, the method development was focused on achieving
Conclusions
The presented method describes a simple, sensitive, and fast assay for seven steroid metabolites in human serum. Unlike methods employing column-switching techniques [24], [27], [30], the chromatographic separation was achieved using a robust setup of single column and HPLC pump and the ionization was carried out with a conventional ESI source. Contrary to the common practice involving several sample preparation steps, the entire process of extraction, derivatization, and injection was
Acknowledgements
This work was supported by the Finnish funding agency for technology and innovation (Tekes). Dr Tiia Kuuranne is acknowledged for valuable advice in determining the method selectivity; Marta Costa and Miia Sinko for their assistance during method development.
References (36)
- et al.
Measurement of steroid sex hormones in serum: a comparison of radioimmunoassay and mass spectrometry
Steroids
(2002) - et al.
Preclinical challenges in steroid analysis of human samples
J. Steroid Biochem. Mol. Biol.
(2010) - et al.
Liquid chromatography tandem mass spectrometry for analysis of steroids in clinical laboratories
Clin. Biochem.
(2011) Clinical steroid mass spectrometry: a 45-year history culminating in HPLC–MS/MS becoming an essential tool for patient diagnosis
J. Steroid Biochem. Mol. Biol.
(2010)- et al.
Advantages and challenges of mass spectrometry assays for steroid hormones
J. Steroid Biochem. Mol. Biol.
(2010) - et al.
Sensitivity enhancement in liquid chromatography/atmospheric pressure ionization mass spectrometry using derivatization and mobile phase additives
J. Chromatogr. B
(2005) - et al.
Rapid measurement of estrogens and their metabolites in human serum by liquid chromatography–tandem mass spectrometry without derivatization
Clin. Biochem.
(2008) - et al.
Development and validation of a sensitive liquid chromatography–tandem mass spectrometry assay to simultaneously measure androgens and estrogens in serum without derivatization
Clin. Chim. Acta
(2009) - et al.
Simultaneous determination of 12 steroids by isotope dilution liquid chromatography–photospray ionization tandem mass spectrometry
Clin. Chim. Acta
(2006) - et al.
Direct measurement of serum free testosterone by ultrafiltration followed by liquid chromatography tandem mass spectrometry
Clin. Biochem.
(2010)
Comparison of electrospray ionization, atmospheric pressure chemical ionization and atmospheric pressure photoionization for determining estrogenic chemicals in water by liquid chromatography tandem mass spectrometry with chemical derivatizations
J Chromatogr. A
Clinical determination of 17-hydroxyprogesterone in serum by LC–MS/MS: comparison to Coat-A-Count RIA method
J. Chromatogr. B
Automated, fast and sensitive quantification of 17 alpha-hydroxy-progesterone, androstenedione and testosterone by tandem mass spectrometry with on-line extraction
Steroids
Low serum testosterone assayed by liquid chromatography–tandem mass spectrometry. Comparison with five immunoassay techniques
Clin. Chim. Acta
Validation of a high throughput method for serum/plasma testosterone using liquid chromatography tandem mass spectrometry (LC–MS/MS)
Steroids
Simultaneous quantitation of seven endogenous C-21 adrenal steroids by liquid chromatography tandem mass spectrometry in human serum
J. Chromatogr. B
Fast and direct quantification of adrenal steroids by tandem mass spectrometry in serum and dried blood spots
J. Chromatogr. B
Validation of a total testosterone assay using high-turbulence liquid chromatography tandem mass spectrometry: total and free testosterone reference ranges
Steroids
Cited by (101)
Validity of the association between five steroid hormones quantification and female infertility conditions: A new perspective for clinical diagnosis
2022, SteroidsCitation Excerpt :Quantitative determination of endogenous sex hormones is important in assessing ovarian function, follicular development and oocyte maturity of female infertility. These sex hormones, such as estradiol (E2), progesterone (P4), and 17-hydroxyprogesterone (17-OHP4), 11-deoxycorticosterone (DOC), and pregnenolone (P5), are important steroid hormones and chemical messengers at low concentrations, which are synthesized from cholesterol via following complex metabolic pathways, as shown in Fig. 1 [5–7]. Depending on the studies, those the main steroid hormones, including E2, P4, and 17-OHP4, play essential roles in the maintenance of ovarian function and sexual characteristics [6,8].
Solid-phase analytical derivatization as a tool for the quantification of steroid hormones in human urine with HPLC-Q-ToF detection
2022, Journal of Pharmaceutical and Biomedical AnalysisCitation Excerpt :To increase the sensitivity of ESI-MS, chemical derivatization of steroids can also be carried out after the SPE procedure, e.g., by reaction of carbonyl groups in steroids with hydroxylamine resulting in the formation of oxime derivatives [6,16]. This derivatization reaction is relatively simple as required only thermostating for 30 min at 60 °C, and it allows direct injection of the obtained products into the chromatographic column [2,23]. Oxime formation improves ionization degree and via this increases the sensitivity of oxosteroids ESI-MS determination by approximately 4–200 times depending on the analyte [24,25].
A review on analysis of steroid profile in different biological matrices
2022, Microchemical Journal