Reduction in MRSA environmental contamination with a portable HEPA-filtration unit

https://doi.org/10.1016/j.jhin.2005.11.011Get rights and content

Summary

There is renewed interest in the hospital environment as a potentially important factor for cross-infection with methicillin-resistant Staphylococcus aureus (MRSA) and other nosocomial pathogens. The aim of this study was to evaluate the effectiveness of a portable high-efficiency particulate air (HEPA)-filtration unit (IQAir Cleanroom H13, Incen AG, Goldach, Switzerland) at reducing MRSA environmental surface contamination within a clinical setting. The MRSA contamination rate on horizontal surfaces was assessed with agar settle plates in ward side-rooms of three patients who were heavy MRSA dispersers. Contamination rates were measured at different air filtration rates (60–235 m3/h) and compared with no air filtration using Poisson regression. Without air filtration, between 80% and 100% of settle plates were positive for MRSA, with the mean number of MRSA colony-forming units (cfu)/10-h exposure/plate ranging from 4.1 to 27.7. Air filtration at a rate of 140 m3/h (one patient) and 235 m3/h (two patients), resulted in a highly significant decrease in contamination rates compared with no air filtration (adjusted rate ratios 0.037, 0.099 and 0.248, respectively; P < 0.001 for each). A strong association was demonstrated between the rate of air filtration and the mean number of MRSA cfu/10-h exposure/plate (P for trend < 0.001). In conclusion, this portable HEPA-filtration unit can significantly reduce MRSA environmental contamination within patient isolation rooms, and this may prove to be a useful addition to existing MRSA infection control measures.

Introduction

Handwashing by healthcare workers is thought to be the most important method of control of methicillin-resistant Staphylococcus aureus (MRSA) and other nosocomial infections.1 However, there is renewed interest in the hospital environment as a potentially important factor in cross-infection.2, 3, 4 The authors recently investigated an outbreak of Hickman-line-associated MRSA bacteraemia in a haematology day-case unit. As part of the investigations, significant MRSA contamination of the environment was found, and settle plates were positive for MRSA on a number of occasions, suggesting ongoing dispersal of MRSA into the air. At least two patients were identified as heavy shedders of MRSA, and the outbreak was controlled when a single room was designated for their further day-case treatment.

This outbreak prompted an investigation of technologies that may reduce future MRSA environmental contamination within such a setting. A number of air decontamination products have been reviewed recently by the Rapid Review Panel of the Health Protection Agency, including air filtration units and ultra-violet air disinfection units (www.hpa.org.uk/infections/topics_az/rapid_review). Two of these products have been recommended for in-use evaluations in a clinical setting (recommendation level 2), including the IQAir Cleanroom H13 portable air purification unit (Incen AG, Goldach, Switzerland). This article presents an evaluation of the effectiveness of the IQAir unit at preventing contamination of environmental surfaces with MRSA in three different clinical settings.

Section snippets

IQAir Cleanroom unit

Two IQAir Cleanroom H13 units were kindly provided by Air Science Ltd (Stroud, UK). The machines were free-standing floor units operating in recirculation mode. This portable air filtration purifier removes airborne micro-organisms by repeatedly drawing ambient room air through a pre-filter and then through a high-efficiency particulate air (HEPA) filter, with an efficiency of 99.97% for particles greater than or equal to 0.3 μm. The IQAir H13 unit's centrifugal fan can be set to run at five

Effect of air filtration on MRSA environmental contamination

Without air filtration, all three patients were confirmed to be heavy dispersers of MRSA (see Table I and Figure 1). Overall, between 80% and 100% of the settle plates were positive, with the mean number of MRSA cfu/10-h exposure/plate ranging from 4.1 to 27.7. The extent and distance of MRSA shedding without air filtration for each patient is shown graphically in Figure 1.

Air filtration at a rate of 235 m3/h (Patients A and C) and 140 m3/h (Patient B) resulted in a significant decrease in MRSA

Discussion

It is well established that MRSA contamination of the environment occurs in the vicinity of MRSA patients, and this is the principal reason for isolating patients in single rooms.5 However, relatively few reported studies have used agar settle (sedimentation) plates to assess the rate of MRSA contamination of environmental surfaces.6 This simple methodology was used in the present study to identify heavy shedders of MRSA and to measure the effect of using a portable HEPA-filtration unit on the

Acknowledgements

The authors are grateful to Air Science Ltd for providing two IQAir H13 Cleanroom units for this study. The authors would also like to thank Dr Sarah Lewis for helping with the statistical analysis.