Continuous non-invasive blood pressure monitoring using concentrically interlocking control loops

https://doi.org/10.1016/j.compbiomed.2005.04.003Get rights and content

Abstract

A new method and apparatus for non-disruptive blood pressure (BP) recording in the finger based on the vascular unloading technique is introduced. The instrument, in contrast to intermittent set point readjustments of the conventional vascular unloading technique, delivers BP without interruptions, thus refining the Peňáz’ principle. The method is based on concentrically interlocking control loops for correct long-term tracing of finger BP, including automatic set point adaptation, light control and separate inlet and outlet valves for electro-pneumatic control. Examples of long-term BP recordings at rest and during autonomic function tests illustrate the potential of the new instrument.

Introduction

For generations, researchers and scientists have been engaged with the non-invasive registration of beat-to-beat arterial blood pressure (BP). The basic principle of vascular unloading dates back to Marey in 1878 [1]. In 1942, Wagner introduced a mechanical system for the registration of BP at the A. radialis using this principle [2], and in 1973, a method for assessment of non-invasive BP by means of an electro-pneumatic control loop was introduced by Peňáz [3], [4]. At that time the Peňáz’ principle enabled the first, but only short, recording of non-invasive beat-to-beat BP using only a single control loop. This control loop is responsible for the fast tracking of BP changes as well as for the stability of the system. However, the changes in arterial diameter and wall tension due to vasoconstriction and vasodilatation render long-term measurements with this single control loop almost impossible, since the true unloading of the arterial wall (the so-called “set point”) is easily lost. Therefore, the Peňáz’ principle has been improved by several groups (see, e.g., [5], [6], [7], [8], [9], [10], [11], [12], [13], [14], [15], [16], [17], [18], [19], [20]), but all of them still use one single control loop which necessitates regular interruptions of the BP tracings for set point recalibration purposes. This recalibration is achieved by opening the servo-control feedback loop and performing a pressure ramp in open-loop mode to recover the set point. This process is clearly a disadvantage of the vascular unloading principle because of data loss during the time of recalibration. Despite this short-coming, the vascular unloading technique has been used extensively and successfully for non-invasive measurement of beat-to-beat blood pressure (see, e.g., [21], [22], [23], [24], [25]).

The present paper describes a method for continuous non-invasive blood pressure measurement (contBP) which no longer needs “open-loop” recalibration of the set point. The contBP module is part of the Task Force® Monitor (TFM), a commercially available monitoring device combining contBP, beat-to-beat stroke volume measurement by impedance cardiography (ICG) and 4 lead ECG [26], [27], [28]. As explained below, contBP is measured on the proximal limb of the index or middle finger by an improved version of the vascular unloading principle using several concentrically interlocking control loops which enhance the accuracy and stability of BP measurement. Each control loop has its own well-defined part within the overall control mechanism: the inner control loops are responsible for fast adjustments and provide near-ideal conditions for the outer control loops which are responsible for the long-term stability of the system.

Due to this new multi-loop control system, truly continuous BP recordings (i.e., without interruptions for recalibration purposes) are possible for the first time. It not only ensures correct long-term BP measurements as will be shown in exemplary comparisons to the Finapres device and simultaneously recorded intra-arterial BP, but also allows the monitoring of beat-to-beat BP during marked rapid changes of blood pressure as observed, e.g., during autonomic function tests. The purpose of the present paper is, therefore, not a systematic comparison with other non-invasive BP recorders or intra-arterial recordings, which will be the content of a second paper, but a description of the technical details of the new method and a short overview of potential applications.

Section snippets

The Peňáz principle revisited

In principle, the Peňáz method with its single electro-pneumatic control loop (see Fig. 1) works as follows: an extremity of the human body containing an artery (e.g., finger, carpus, temple, etc.) is illuminated with infrared (IR) light. Part of the light is absorbed by the pulsatile blood volume so that the reflected light is an inverse measure of the pulsatile blood volume. Vendrik and Vierhout [29] concluded that a finger is the most suitable place for this method, but the accuracy would be

Results

A systematic comparison of the new method with intra-arterial blood pressure is the subject of a separate paper. In the present paper, the high-quality performance of the new instrument for the vascular unloading technique is shown by demonstrating:

  • (a)

    Exemplary comparisons with the currently only other commercially available instrument for vascular unloading at the finger, the Finapres device.

  • (b)

    Examples of intra-arterial recordings as compared to the present technology of vascular unloading and

  • (c)

    the

Discussion

As described in the present paper, a number of major improvements of Peňáz's vascular unloading technique for long-term non-invasive BP recordings have been achieved. The use of several interlocking control loops arranged concentrically according to the necessity for speed (inner loops) or long-term optimization (outer loops) for the first time enables truly continuous non-interrupted non-invasive blood pressure measurement over long periods of time. The automatic detection and correction of

Summary

The present paper describes improvements for continuous non-invasive blood pressure (BP) measurement based on the vascular unloading technique which no longer necessitate “open-loop” calibration readjustment of the set point and is therefore the first method to measure non-disruptive, truly continuous finger BP. The so-called contBP module is part of the Task Force® Monitor (TFM), a commercially available monitoring device combining contBP, beat-to-beat stroke volume measurement by impedance

Acknowledgements

This work was supported by the Austrian Science Fund FWF, Sonderforschungsbereich Biomembranes 007, by the Austrian Industrial Research Promotion Fund (FFF), by the Styrian Business Promotion Agency (SFG) and CNSystems Medizintechnik GmbH. The authors want to thank Dr. Doris Flotzinger, Priv. Doz. Dr. Michael Imhoff and Derrick Ebden for their critical review of the paper.

Jürgen Fortin received his MSc (Dipl.-Ing.) in Biomedical Engineering for work on Impedance Cardiography from the Technical University Graz, Austria, in 1995 where he is currently working for a PhD in Medical Engineering. In 1998, he cofounded and became CEO of an Austria-based company which develops non-invasive cardiovascular monitors and has received several national and international innovation awards. His main research interests are the regulation mechanisms of hemodynamics— “the heart and

References (34)

  • N.T. Smith et al.

    Evaluation of two prototype devices producing non-invasive, pulsatile, calibrated blood pressure measurement from a finger

    J. Clin. Monit.

    (1985)
  • T. Kurki et al.

    Non-invasive continuous blood pressure measurement from the finger: optimal measurement conditions and factor affecting reliability

    J. Clin. Monit.

    (1987)
  • B.P.M. Imholz et al.

    Continuous non-invasive blood pressure monitoring: reliability of Finapres during the Valsalva manoeuvre

    Cardiovasc. Res.

    (1988)
  • B.P.M. Imholz et al.

    Non-invasive beat-to-beat finger blood pressure measurement during orthostatic stress compared to intra-arterial pressure

    Cardiovasc. Res.

    (1990)
  • M. Nakagawara et al.

    A portable instrument for non-invasive monitoring of beat-by-beat cardiovascular hemodynamic parameters based on the volume-compensation and electrical-admittance method

    Med. Biol. Eng. Comput.

    (2000)
  • K. Yamakoshi et al.

    Indirect measurement of instantaneous arterial blood pressure in the human finger by the vascular unloading technique

    IEEE Trans. Biomed. Eng.

    (1980)
  • K. Yamakoshi et al.

    Noninvasive measurement of arterial blood pressure and elastic properties using photoelectric plethysmography technique

    Med. Progress Technol.

    (1987)
  • Cited by (164)

    • Sex differences in the age-related decrease of spontaneous baroreflex function in healthy individuals

      2024, American Journal of Physiology - Heart and Circulatory Physiology
    View all citing articles on Scopus

    Jürgen Fortin received his MSc (Dipl.-Ing.) in Biomedical Engineering for work on Impedance Cardiography from the Technical University Graz, Austria, in 1995 where he is currently working for a PhD in Medical Engineering. In 1998, he cofounded and became CEO of an Austria-based company which develops non-invasive cardiovascular monitors and has received several national and international innovation awards. His main research interests are the regulation mechanisms of hemodynamics— “the heart and its control”.

    Wolfgang Marte received his M.D. degree from the Karl-Franzens University Graz, Austria, School of Medicine in 1986. He was a post-doctoral fellow and resident at the Department of Anesthesiology and Intensive Care Medicine at the University Hospital of Graz from 1990 to 2001. Currently, he is Assistant Professor of Anesthesiology and Intensive Care Medicine at the Medical University of Graz, Austria. His research interests include implementation of non-invasive hemodynamic monitoring in the postoperative setting, cardiac rechronisation by MsS (Multisite cardiac stimulation) and implementation of POCT-Technology in ICU.

    Rupert Grüllenberger received his MSc (Dipl.-Ing.) in Electro- and Biomedical Engineering at the University of Technology Graz, Austria, in January 1999. His master thesis focussed on researching and developing tools for measuring parameters like continuous blood pressure, impedance cardiography and electrocardiography. He is currently working on a PhD in Medical Engineering with his research interests including the full range of cardiovascular parameters.

    Alexander Hacker studied Electronics and Communications Engineering at the Technical University in Graz, Austria, working at the first prototypes of the Task Force® Monitor. His research interests include advanced signal processing and RF electronics.

    Walter Habenbacher received his MSc (Dipl.-Ing.) in Biomedical Engineering from the Technical University Graz, Austria, in 1998. Currently, he is responsible for software development in a company developing non-invasive hemodynamic monitors and is working towards a PhD in Medical Engineering at the Technical University Graz. His main research interests are heart rate and blood pressure variability.

    Arnulf Heller graduated from the Graz University of Technology in 2000. He has earned his MSc (Dipl.-Ing.) degree in Telematics and is working towards a PhD in Biomedical Engineering at the Graz University of Technology, Austria. He is currently chief software developer with a company that develops non-invasive medical technologies. His research interests are biomedical signal processing and pulse wave analysis.

    Christian Wagner graduated from FH Technikum Kaernten, Austria, receiving his MSc (DI FH) degree in Electronics in 1999. He is currently responsible for quality management and international regulatory affairs with a company that develops non-invasive medical products.

    Paul Wach received his PhD in Electrical Engineering from Graz University of Technology in 1967. He was a fellow at the Institute of Electrical Power Systems in 1966, at the newly founded Institute of Biomedical Engineering from 1973 to 2004 and at the Institute of Medical Engineering since 2004. Currently he is University Professor, head of the Institute of Medical Engineering at Graz, University of Technology, Austria. His research activities include network theory, biological control, modeling and measurement, biophysics of glucose metabolism and the electrical heart function (electric and magnetic heart fields) and inverse problems in non-invasive functional imaging for medical diagnosis.

    Falko Skrabal earned his M.D. from Karl-Franzens University Graz in 1967. He was a post-doctoral fellow at the Royal Postgraduate Medical School in London from 1968 to 1969. He worked at the Medical University Innsbruck from 1970 to 1987; Diploma in Internal Medicine in 1974, Lecturer in Internal Medicine in 1979, Election to Prof. of Endocrinology at the University Innsbruck Medical School in 1986. Further he obtained diplomas in Laboratory Medicine and Clinical Chemistry (1985) and Nephrology (1987). Since 1988 he is head of the Department of Internal Medicine, since 2000 Medical Director, Krankenhaus Barmherzige Brueder, Teaching Hospital Medical University Graz, Austria. His main interests include short-term and long-term blood pressure regulation, diabetes and osteoporosis.

    View full text