Reliability of smartphone measurements of vital parameters: A prospective study using a reference method

https://doi.org/10.1016/j.ajem.2019.03.021Get rights and content

Abstract

Objective

In this study, we aimed to evaluate the accuracy of HR and SaO2 data obtained using a smartphone compared with the measurements of a vital signs monitor (VSM) and an arterial blood gas (ABG) device, respectively.

Material and methods

In this single-center prospective study, the HR and SaO2 measurements were performed using the built-in sensor and light source of a Samsung Galaxy S8 smartphone and compared to the results of VSM and ABG device. The Bland-Altman analysis was used to evaluate and visualize the agreement between the methods.

Results

The data of 101 patients were analyzed. There was a high correlation between HR measured by smartphone and HR measured by VSM [P < 0.0001; 0.9918 (95% CI = 0.987–0.994)]. In addition, the SaO2 values obtained by smartphone were highly correlated with those by ABG (P < 0.0001; 0.968 (95% CI = 0.952–0.978)).

Conclusion

The HR and SaO2 values obtained by smartphone were found to be consistent with the measurements of the reference devices. With the growing use of smartphone technology in the health field, we foresee that patients will be able to make their own triage assessment before presenting to the hospital.

Introduction

Heart rate (HR) and oxygen saturation (SaO2) are two basic parameters in the evaluation of patients admitted to the emergency department. In certain cases, having prior knowledge of these parameters may be helpful in referring patients to the hospital and making an accurate and timely diagnosis. Conventionally, HR is measured by manual or mechanical palpation of peripheral arterial pulses [1]. However, palpation of pulses or SaO2 measurement may be difficult to undertake for patients or their relatives that do not have the necessary medical training or skills. As a result, some patients apply to health centers simply for the determination of their vital parameters [2].

Today, there are many free software applications utilizing the processing capabilities of smartphones. These applications are specifically designed for self-use and allow users to obtain HR in minutes (bpm) and SaO2 levels in %. In recent years, with the addition of certain devices and software to smartphones, non-medically trained individuals can also measure parameters, such as HR and SaO2 at any time and place. This allows patients with dyspnea, blood pressure disorder, and heart rhythm to obtain information about their own vital parameters through a smartphone that is widely available at any time, without requiring a specific device or incurring any additional cost.

The assessment and follow-up of vital signs undertaken by mobile devices not only allow emergency medical intervention to be undertaken at an early stage when there is a deterioration in his/her health status, but also reduces unnecessary hospital admissions, time spent on transport, and health-related costs [3]. Garde et al. argued that health data obtained from mobile devices could be useful in terms of identifying patients that need to be admitted to hospital [4].

Although smartphones are calibrated to perform the measurement of vital parameters, there is a lack of validation studies concerning applications that only use the built-in technology of these devices [5,6]. The measurement of arterial blood gas (ABG) analysis is the gold standard in SaO2 measurement [7]. To the best of our knowledge, to date, no studies have been conducted to compare the ABG results to determine the accuracy of smartphone SaO2 measurements in patients admitted to the emergency department. Therefore, in this study, we aimed to evaluate the accuracy of HR and SaO2 data obtained using a smartphone compared with the measurements of a vital signs monitor (VSM) and an ABG device, respectively.

Section snippets

Material and methods

The study was approved by the University Review Board (dated 22.01.2018 and numbered 2017/KK/150).

Results

The results of 13 of 114 patients were excluded from the study for technical reasons (being unable to verify the ABG results due to drawing venous instead of arterial blood and clotted blood samples). The data of a total of 101 patients, 48 male (47.5%) and 53 female (52.5%), were analyzed. 48 male (47.5%) and 53 female (52.5%) patients were included in the study. The mean age of the male and female patients was 68.08 and 72 years, respectively. According to the age distribution of the

Discussion

Mobile applications are software programs that run on smartphones or other mobile communication devices. These programs may be either accessories added to these devices or a combination of accessories and software. For the appropriate follow-up and diagnosis of patients, it is crucial to determine their vital parameters, such as HR and SaO2. Recently, smartphones have been increasingly used for many purposes in the health field, particularly for the measurement of HR and SaO2.

Smartphone

Abbreviations

    SPSS

    Statistical Package for the Social Science

    SaO2

    oxygen saturation

    HR

    heart rate

    VSM

    vital signs monitor

    ABG

    arterial blood gas

    PPG

    photoplethysmography

    CBA

    camera based application

    PBA

    probe based application

Conflicts of interest

The authors declare that they have no conflicts of interest.

Funding acknowledgement statement

This research received no specific grant from any funding agency in the public, commercial, or not-for-profit sectors.

References (14)

  • W. Meyer-Sabellek et al.

    Non-invasive ambulatory blood pressure monitoring: technical possibilities and problems

    J Hypertens Suppl

    (1990)
  • S. Yıldız et al.

    Evaluating individual characteristics and applications of elderly patients presented to emergency service

    Gazi Sağlık Bilimleri Dergisi

    (2016)
  • R. Wootton

    Twenty years of telemedicine in chronic disease management - an evidence synthesis

    J Telemed Telecare

    (2012)
  • A. Garde et al.

    Respiratory rate and pulse oximetry derived information as predictors of hospital admission in young children in Bangladesh: a prospective observational study

    BMJ Open

    (2016)
  • N.G. Cortez et al.

    FDA regulation of mobile health technologies

    N Engl J Med

    (2014)
  • N. Bruining et al.

    Task force of the e-cardiology working. Acquisition and analysis of cardiovascular signals on smartphones: potential, pitfalls and perspectives: by the task force of the e-cardiology working group of European Society of Cardiology

    Eur J Prev Cardiol

    (2014)
  • E. Zeserson et al.

    Correlation of venous blood gas and pulse oximetry with arterial blood gas in the undifferentiated critically ill patient

    J Intensive Care Med

    (2018)
There are more references available in the full text version of this article.

Cited by (31)

  • Sensor-based techniques for detection of COVID-19

    2021, COVID-19 in the Environment: Impact, Concerns, and Management of Coronavirus
  • Monitoring of heart rate, blood oxygen saturation, and blood pressure using a smartphone

    2020, Biomedical Signal Processing and Control
    Citation Excerpt :

    For home monitoring, devices such as pulse oximeter (PO) or oscillometric sphygmomanometer, which are more user-friendly but usually less accurate, exist. HR can be estimated using one of the following smartphone sensors: rear camera [2–6,34,36], front camera [31,40], accelerometer [38,41], microphone [39], or sensors dedicated to HR and SpO2 measurement [29]. The data from the rear camera are usually recorded from the finger, processed into the photoplethysmogram (PPG) waveform, and further analyzed to calculate HR [2–6,34,36].

View all citing articles on Scopus
View full text