PT - JOURNAL ARTICLE AU - Marek Jastrzębski AU - Grzegorz Kiełbasa AU - Paweł Moskal AU - Agnieszka Bednarek AU - Aleksander Kusiak AU - Tomasz Sondej AU - Adam Bednarski AU - Marek Rajzer AU - Pugazhendhi Vijayaraman TI - Fixation beats – a novel marker for reaching the left bundle branch area during deep septal lead implantation AID - 10.1101/2020.10.04.20206573 DP - 2020 Jan 01 TA - medRxiv PG - 2020.10.04.20206573 4099 - http://medrxiv.org/content/early/2020/10/06/2020.10.04.20206573.short 4100 - http://medrxiv.org/content/early/2020/10/06/2020.10.04.20206573.full AB - Introduction One of the challenges of left bundle branch (LBB) pacing is to place the pacing lead deep enough in the septum to obtain capture of the LBB, yet not too deep to avoid perforation. We hypothesized that the occurrence of the ectopic beats of qR/rsR’ morphology in V1 lead (fixation beats) during the lead fixation would predict that the final desired intraseptal lead depth was just reached, while the lack of fixation beats would indicate too shallow position, and need for more lead rotations.Methods Consecutive patients during LBB pacing device implantation were analyzed retrospectively and then prospectively with respect to the occurrence of the fixation beats during each episode of lead rotation. We compared the presence of fixation beats during the lead rotation event directly before the LBB capture area depth was reached versus during the events before intermediate/unsuccessful positions.Results A total of 339 patients and 1278 lead rotation events were analyzed. In the retrospective phase, the fixation beats were observed in 327/339 of final lead positions and in 9/939 of intermediate lead positions (p<0.001). The sensitivity, specificity, positive and negative predictive value of the LBB area fixation beats as a marker for reaching the LBB capture area was 96.4%, 99.0%, 97.3% and 99.0%, respectively. In the prospective, fixation-beats-guided, implantation phase the fixation beats were observed in all patients and only at the LBB capture depth.Conclusions Monitoring fixation beats during deep septal lead deployment can facilitate the procedure and possibly increase the safety of lead implantation.Competing Interest StatementDr. Vijayaraman: honoraria, consultant, research, fellowship support: Medtronic; consultant Abbott, Biotronik, Boston Scientific, Eaglepoint LLC; patent for His delivery tool Dr. Jastrzebski: consultant and lecture fees from Medtronic. Funding Statementno fundingAuthor DeclarationsI confirm all relevant ethical guidelines have been followed, and any necessary IRB and/or ethics committee approvals have been obtained.YesThe details of the IRB/oversight body that provided approval or exemption for the research described are given below:Jagiellonian University Bioethical CommitteeAll necessary patient/participant consent has been obtained and the appropriate institutional forms have been archived.YesI understand that all clinical trials and any other prospective interventional studies must be registered with an ICMJE-approved registry, such as ClinicalTrials.gov. I confirm that any such study reported in the manuscript has been registered and the trial registration ID is provided (note: if posting a prospective study registered retrospectively, please provide a statement in the trial ID field explaining why the study was not registered in advance).YesI have followed all appropriate research reporting guidelines and uploaded the relevant EQUATOR Network research reporting checklist(s) and other pertinent material as supplementary files, if applicable.YesData available upon reasonable request