PT - JOURNAL ARTICLE AU - I Sayegh, Samir TI - Consequences of Mismatch, Misalignment and Rotation of Toric Intraocular Lenses in Refractive Cataract Surgery Part 2. Avoiding Flip Flops AID - 10.1101/2020.09.30.20203380 DP - 2020 Jan 01 TA - medRxiv PG - 2020.09.30.20203380 4099 - http://medrxiv.org/content/early/2020/09/30/2020.09.30.20203380.short 4100 - http://medrxiv.org/content/early/2020/09/30/2020.09.30.20203380.full AB - Purpose To show current approaches for overcorrecting astigmatism and “flipping” its axis need be reconsidered in light of methods we introduce that take into account both mismatch and misalignment of the toric intraocular lens (tIOL) with respect to the astigmatism to be corrected at the time of cataract surgery.Setting Private Practice and Research Center. The EYE Center. Champaign, IL, USA.Design Formal Analytical StudyMethods In the most common surgical situation where both mismatch and misalignment exist, we present an analysis of the point at which overcorrection and undercorrection residuals coincide, yielding a simple but powerful methodology to predict the optimal degree of overcorrection with a tIOL. The method is illustrated for tIOLs used in surgical practice.Results The minimum astigmatism appropriate to overcorrect with a tIOL is given by, ,where m is the midpoint threshold used by “split-the-difference” algorithms and χ is the estimate of tIOL misalignment due to all causes. Correspondingly, the maximum overcorrection, Ωmax, that should be attempted is where is the dioptric step at the corneal plane, with σ = H − B, where H = n σ is the cylinder of the overcorrecting tIOL and B = (n − 1)σ is the cylinder of the undercorrecting tIOL, both at the IOL plane, τis the toricity ratio and γ relates to the angle of misalignment χ by which can be approximated by . Ω maxfactors elegantly in the product of , the (ideal) midpoint correction for perfect alignment, by the bracketed term, representing the percent reduction of the ideal value in a realistic surgical situation with estimated misalignment χ. To illustrate: an eye of average dimensions and tIOLs from major manufacturers , with A = 2.35 D dictating n = 5. For a misalignment of 10°Ωmax≅ 0.10 D is the maximum overcorrection that should be accepted, significantly smaller than the midpoint D, recommended by many current tIOL calculators.Conclusion An optimal method is presented for the selection of an overcorrecting tIOL at the time of refractive cataract surgery with improvement over current tIOL calculators’ methods.Competing Interest StatementThe authors have declared no competing interest.Funding StatementNo external funding was received.Author DeclarationsI confirm all relevant ethical guidelines have been followed, and any necessary IRB and/or ethics committee approvals have been obtained.YesThe details of the IRB/oversight body that provided approval or exemption for the research described are given below:No IRB/oversight body was required in preparing this manuscript.All necessary patient/participant consent has been obtained and the appropriate institutional forms have been archived.YesI understand that all clinical trials and any other prospective interventional studies must be registered with an ICMJE-approved registry, such as ClinicalTrials.gov. I confirm that any such study reported in the manuscript has been registered and the trial registration ID is provided (note: if posting a prospective study registered retrospectively, please provide a statement in the trial ID field explaining why the study was not registered in advance).YesI have followed all appropriate research reporting guidelines and uploaded the relevant EQUATOR Network research reporting checklist(s) and other pertinent material as supplementary files, if applicable.YesAll necessary data exists within the manuscript.