PT - JOURNAL ARTICLE AU - van Oosten, Julien P. AU - Goedendorp, Nico AU - Mousa, Amne AU - Flink, Rutger AU - Schaart, Rik AU - Flinsenberg, Merel AU - Somhorst, Peter AU - Gommers, Diederik A.M.P.J. AU - Heunks, Leo AU - Jonkman, Annemijn H. TI - Solid-state esophageal pressure sensor for the estimation of pleural pressure: a bench and first-in-human validation study AID - 10.1101/2024.10.01.24314687 DP - 2024 Jan 01 TA - medRxiv PG - 2024.10.01.24314687 4099 - http://medrxiv.org/content/early/2024/10/01/2024.10.01.24314687.short 4100 - http://medrxiv.org/content/early/2024/10/01/2024.10.01.24314687.full AB - Background Advanced respiratory monitoring through the measurement of esophageal pressure (Pes) as a surrogate of pleural pressure helps guiding mechanical ventilation in ICU patients. Pes measurement with an esophageal balloon catheter, the current clinical reference standard, needs complex calibrations and a multitude of factors influence its reliability. Solid-state pressure sensors might be able to overcome these limitations.Objectives To evaluate the accuracy of a new solid-state Pes transducer (Pessolid). We hypothesized that measurements are non-inferior to those obtained with a properly calibrated balloon catheter (Pesbal).Methods Absolute and relative solid-state sensor Pes measurements were compared to a reference pressure in a 5-day bench setup, and to simultaneously placed balloon catheters in 15 spontaneously breathing healthy volunteers and in 16 mechanically ventilated ICU patients. Bland-Altman analysis was performed with nonparametric bootstrapping to estimate bias and upper and lower limits of agreement (LoA).Results Bench study: Solid-state pressure transducers had a positive bias (Psolid – Pref) of around 1 cmH2O for the absolute minimal and maximum pressures, and no bias for pressure swings. Healthy volunteers: the solid-state transducer revealed a bias (Pessolid–Pesbal) [upper LoA; lower LoA] of 1.58 [8.19; −5.03], −2.37 [3.96; −8.69] and 3.94 [11.09; −3.20] cmH2O for end-expiratory, end-inspiratory and ΔPes values, respectively. ICU patients: the solid-state transducer showed a bias (Pessolid–Pesbal) [upper LoA; lower LoA] during controlled / assisted ventilation of: −0.15 [1.39; −1.70] / −0.20 [5.02; −5.41], 0.32 [3.35; −2.72] / −0.54 [4.60; −5.68] and 0.47 [3.79; −2.85] / 0.35 [3.88; −3.18] cmH2O for end-expiratory, end-inspiratory and ΔPes values, respectively. LoA were <2cmH2O for static measurements on controlled ventilation.Conclusions The novel solid-state pressure transducer showed good accuracy on the bench, in healthy volunteers and in ventilated ICU-patients. This could contribute to the implementation of Pes as advanced respiratory monitoring technique.Trial registration Clinicaltrials.gov identifier: NCT05817968 (patient study). Registered on 18 April 2023.Competing Interest StatementRF and RS are employed by Pulmotech B.V. AJ and LH received research funding (paid to the institution) of Pulmotech B.V. for the conduct of the study.Clinical TrialNCT05817968Funding StatementThe study was partially funded by a grant from Pulmotech B.V., Leek, The Netherlands.Author DeclarationsI confirm all relevant ethical guidelines have been followed, and any necessary IRB and/or ethics committee approvals have been obtained.YesThe details of the IRB/oversight body that provided approval or exemption for the research described are given below:This study was approved by the local medical ethics committee of the Vrije Universiteit medical center (METc 2020.470) and Erasmus Medical Center (MEC-2023-0119). All healthy subjects and patients provided written informed consent prior to participation.I confirm that all necessary patient/participant consent has been obtained and the appropriate institutional forms have been archived, and that any patient/participant/sample identifiers included were not known to anyone (e.g., hospital staff, patients or participants themselves) outside the research group so cannot be used to identify individuals.YesI understand that all clinical trials and any other prospective interventional studies must be registered with an ICMJE-approved registry, such as ClinicalTrials.gov. I confirm that any such study reported in the manuscript has been registered and the trial registration ID is provided (note: if posting a prospective study registered retrospectively, please provide a statement in the trial ID field explaining why the study was not registered in advance).YesI have followed all appropriate research reporting guidelines, such as any relevant EQUATOR Network research reporting checklist(s) and other pertinent material, if applicable.YesAll data produced in the present study are available upon reasonable request to the authorsBMIBody Mass IndexICUIntensive Care UnitiEPCIntelligent Esophageal Pressure CatheteriEPMSIntelligent Esophageal Pressure Monitoring SystemLoALimit of AgreementPawAirway pressurePesEsophageal pressurePesbalEsophageal pressure measured by balloon catheterPrefReference pressurePsolidEsophageal pressure measured by the solid-state pressure sensor