TY - JOUR T1 - Closed-loop control of k-space sampling via physiologic feedback for cine MRI JF - medRxiv DO - 10.1101/2020.06.22.20137638 SP - 2020.06.22.20137638 AU - Francisco Contijoch AU - Yuchi Han AU - Srikant Kamesh Iyer AU - Peter Kellman AU - Gene Gualtieri AU - Mark A. Elliott AU - Sebastian Berisha AU - Joseph H. Gorman III AU - Robert C. Gorman AU - James J. Pilla AU - Walter R. Witschey Y1 - 2020/01/01 UR - http://medrxiv.org/content/early/2020/06/23/2020.06.22.20137638.abstract N2 - Background Segmented cine cardiac MRI combines data from multiple heartbeats to achieve high spatiotemporal resolution cardiac images, yet predefined k-space segmentation trajectories can lead to suboptimal k-space sampling. In this work, we developed and evaluated an autonomous and closed-loop control system for radial k-space sampling to increase sampling uniformity.Methods The closed-loop system autonomously selects radial k-space sampling trajectory during live segmented cine MRI and attempts to optimize angular sampling uniformity by selecting views in regions of k-space that were not previously well-sampled. Sampling uniformity and robustness to arrhythmias was assessed using ECG data acquired from 10 normal subjects in an MRI scanner. The approach was then implemented with a fast gradient echo sequence on a whole-body clinical MRI scanner and imaging was performed in 4 healthy volunteers. The closed-loop k-space trajectory was compared to random, uniformly distributed and golden angle view trajectories via measurement of k-space uniformity and the point spread function. Lastly, an arrhythmic dataset was used to evaluate a potential application of the approach.Results The autonomous trajectory increased k-space sampling uniformity by 13±7%, main lobe point spread function (PSF) signal intensity by 14±6%, and reduced ringing relative to golden angle sampling. When implemented, the autonomous pulse sequence prescribed radial view angles faster than the scan TR (0.98 ± 0.02 ms, maximum = 1.38 ms) and increased k-space sampling mean uniformity by 5±12%, decreased uniformity variability by 45±14%, and increased PSF signal ratio by 5±5% relative to golden angle sampling.Conclusion The closed-loop approach enables near-uniform radial sampling in a segmented acquisition approach which was higher than predetermined golden-angle radial sampling. This can be utilized to increase the sampling or decrease the temporal footprint of an acquisition and the closed-loop framework has the potential to be applied to patients with complex heart rhythms.Competing Interest StatementThe authors have declared no competing interest.Funding StatementThis study was supported by the National Institutes of Health Heart, Lung, and Blood Institute. Specifically, FC was supported by HL120580, WRTW was supported by HL108157, and RCG was supported by R01 HL063954. The funders had no role in study design, data collection and analysis, decision to publish, or preparation of the manuscript.Author DeclarationsI confirm all relevant ethical guidelines have been followed, and any necessary IRB and/or ethics committee approvals have been obtained.YesThe details of the IRB/oversight body that provided approval or exemption for the research described are given below:All subjects and patients gave informed consent prior to participating in the study (817383), approved by the Institutional Review Board of the University of Pennsylvania.All necessary patient/participant consent has been obtained and the appropriate institutional forms have been archived.YesI understand that all clinical trials and any other prospective interventional studies must be registered with an ICMJE-approved registry, such as ClinicalTrials.gov. I confirm that any such study reported in the manuscript has been registered and the trial registration ID is provided (note: if posting a prospective study registered retrospectively, please provide a statement in the trial ID field explaining why the study was not registered in advance).YesI have followed all appropriate research reporting guidelines and uploaded the relevant EQUATOR Network research reporting checklist(s) and other pertinent material as supplementary files, if applicable.YesData cannot be shared publicly because of commercial research agreements with imaging vendors. Data are available (contact via Dr Contijoch) for researchers who meet the criteria for access to confidential data.DSPdigital signal processECGelectrocardiogramFOVfield-of-viewRFradiofrequencyPSFpoint spread functionTEecho timeTRrepetition time ER -