RT Journal Article SR Electronic T1 Repeatability of a Dual-Scheimpflug Placido Disc Corneal Tomographer/Topographer in Eyes with Keratoconus JF medRxiv FD Cold Spring Harbor Laboratory Press SP 2020.05.13.20067710 DO 10.1101/2020.05.13.20067710 A1 Poulsen, Anne A1 Jang, Diane A1 Khan, Mahmood A1 Al-Mohtaseb, Zaina Nabil A1 Chen, Michael A1 Zhan, Xiang A1 Banerjee, Kalins A1 Scott, Ingrid U A1 Pantanelli, Seth M YR 2020 UL http://medrxiv.org/content/early/2020/05/18/2020.05.13.20067710.abstract AB Purpose To investigate the repeatability of a combined Dual-Scheimpflug placido disc corneal topographer (Zeimer Galilei G4) with respect to keratometric indices used to monitor progression of keratoconus (KCN).Methods Patients with KCN were prospectively enrolled. For each eye lacking history of corneal surgery, 5 measurements were taken in succession. Eyes in which 3 or more measurements could be obtained (defined by the device’s 4 image quality metrics) were included in the analysis. The repeatability limits (RL) and interclass correlation coefficients (ICC) were calculated for various parameters.Results 32 eyes from 25 patients met all image quality metrics, and 54 eyes from 38 patients met at least 3/4 criteria (all except the placido image quality metric). RLs for key parameters when 4/4 or ≥3/4 image quality metrics were met included: 0.37 and 0.77 diopters (D) for steep simulated keratometry, 0.79 and 1.65 D for maximum keratometry, 13.80 and 13.88 degrees for astigmatism axis, 0.64 and 0.56 μm for vertical coma magnitude, and 3.76 and 3.84 μm for thinnest pachymetry, respectively. The ICCs for all parameters were excellent [above 0.87 except for spherical aberration (0.77)].Conclusions The dual-Scheimpflug placido disc corneal topographer is highly repeatable in quantifying parameters used in monitoring KCN. Excellent placido images are difficult to capture in eyes with KCN, but when available, increase the reliability of the measurements. The RLs may be especially helpful in detecting progression in mild KCN when interventions such as corneal cross-linking or intrastromal corneal ring segments are most beneficial.Competing Interest StatementThe authors have declared no competing interest.Clinical TrialNCT03511495Funding StatementNo external funding was received in relation to this project.Author DeclarationsAll relevant ethical guidelines have been followed; any necessary IRB and/or ethics committee approvals have been obtained and details of the IRB/oversight body are included in the manuscript.YesAll necessary patient/participant consent has been obtained and the appropriate institutional forms have been archived.YesI understand that all clinical trials and any other prospective interventional studies must be registered with an ICMJE-approved registry, such as ClinicalTrials.gov. I confirm that any such study reported in the manuscript has been registered and the trial registration ID is provided (note: if posting a prospective study registered retrospectively, please provide a statement in the trial ID field explaining why the study was not registered in advance).YesI have followed all appropriate research reporting guidelines and uploaded the relevant EQUATOR Network research reporting checklist(s) and other pertinent material as supplementary files, if applicable.YesFull dataset may be available to interested parties by request. Please e-mail the corresponding author to query.