PT - JOURNAL ARTICLE AU - Plummer, Andrew AU - du Bois, Jonathan L. AU - Flynn, Joseph AU - Roesner, Jens AU - Lee, Siu Man AU - Magee, Patrick AU - Thornton, Malcolm AU - Padkin, Andrew AU - Gill, Harinderjit S. TI - A Simple Method to Estimate Flow Restriction for Dual Ventilation of Dissimilar Patients: The BathRC Model AID - 10.1101/2020.04.12.20062497 DP - 2020 Jan 01 TA - medRxiv PG - 2020.04.12.20062497 4099 - http://medrxiv.org/content/early/2020/05/15/2020.04.12.20062497.short 4100 - http://medrxiv.org/content/early/2020/05/15/2020.04.12.20062497.full AB - Objectives With large numbers of COVID-19 patients requiring mechanical ventilation and ventilators possibly being in short supply, in extremis two patients may have to share one ventilator. Careful matching of patient ventilation requirements is necessary. However, good matching is difficult to achieve as lung characteristics can have a wide range and may vary over time. Adding flow restriction to the flow path between ventilator and patient gives the opportunity to control the airway pressure and hence flow and volume individually for each patient. This study aimed to create and validate a simple model for calculating required flow restriction.Design/Setting Experimental/Secondary CareMethods We created a simple linear resistance-compliance model, termed the BathRC model, of the ventilator tubing system and lung allowing direct calculation of the relationships between pressures, volumes, and required flow restriction. Experimental measurements were made for parameter determination and validation using a clinical ventilator connected to two test lungs. For validation, differing amounts of restriction were introduced into the ventilator circuit.Results The BathRC model was able to predict tidal lung volumes with a mean error of 4% (min:1.2%, max:9.3%).Conclusion We present a simple model validated model that can be used to estimate required flow restriction for dual patient ventilation. The BathRC model is freely available; this tool is provided to demonstrate that flow restriction can be readily estimated.ARTICLE SUMMARY Strengths and limitations of this studyData collected on a clinical systemSimple model, easy to implementValidation performedLimited to 2 types of restrictors testedData from only one ventilator typeCompeting Interest StatementThe authors have declared no competing interest.Funding StatementNo external funding.Author DeclarationsAll relevant ethical guidelines have been followed; any necessary IRB and/or ethics committee approvals have been obtained and details of the IRB/oversight body are included in the manuscript.YesAll necessary patient/participant consent has been obtained and the appropriate institutional forms have been archived.YesI understand that all clinical trials and any other prospective interventional studies must be registered with an ICMJE-approved registry, such as ClinicalTrials.gov. I confirm that any such study reported in the manuscript has been registered and the trial registration ID is provided (note: if posting a prospective study registered retrospectively, please provide a statement in the trial ID field explaining why the study was not registered in advance).YesI have followed all appropriate research reporting guidelines and uploaded the relevant EQUATOR Network research reporting checklist(s) and other pertinent material as supplementary files, if applicable.YesBathRC model is available here for general usage. https://github.com/arp23/BathRC