TY - JOUR T1 - Evaluation of a rapid lateral flow calprotectin test for the diagnosis of prosthetic joint infection JF - medRxiv DO - 10.1101/19004473 SP - 19004473 AU - Alexander J. Trotter AU - Rachael Dean AU - Celia E. Whitehouse AU - Jarle Mikalsen AU - Claire Hill AU - Roxanne Brunton-Sim AU - Gemma L. Kay AU - Majeed Shakokhani AU - Alexander Durst AU - John Wain AU - Iain McNamara AU - Justin O’Grady Y1 - 2019/01/01 UR - http://medrxiv.org/content/early/2019/08/17/19004473.abstract N2 - Background Microbiological diagnosis of prosthetic joint infection (PJI) relies on culture techniques that are slow and insensitive. Rapid tests are urgently required to improve patient management. Calprotectin is a neutrophil biomarker of inflammation that has been demonstrated to be effective for the diagnosis of PJI. A calprotectin based lateral flow test has been developed for the rapid detection of PJI using synovial fluid samples.Methods A convenience series of 69 synovial fluid samples from patients at the Norfolk and Norwich University Hospitals (NNUH) were collected intraoperatively from 52 hip and 17 knee revision operations. Calprotectin levels were measured using a new commercially available lateral flow assay for PJI diagnosis (Lyfstone). For all samples, synovial fluid was pipetted onto the lateral flow device and the signal was read using a mobile phone app after 15 minutes incubation at room temperature.Results According to the Musculoskeletal Infection Society (MSIS) criteria, 24 patients were defined as PJI positive and the remaining 45 were negative. The overall accuracy of the lateral flow test against the MSIS criteria was 75%. The test had a sensitivity and specificity of 75% and 76% respectively with a positive predictive value (PPV) of 62% and a negative predictive value (NPV) of 85%. Discordant results were then reviewed by the clinical team using available patient data to develop an alternative gold standard for defining presence/absence of infection (MSIS+). Compared to MSIS+, the test showed an overall accuracy of 83%, sensitivity and specificity of 95% and 78% respectively, a PPV of 62% and an NPV of 98%. Test accuracy for hip revisions was 77% and for knee revisions was 100%.Conclusions This study demonstrates that the calprotectin lateral flow assay is an effective diagnostic test for PJI. Our data suggests that the test is likely to generate false positive results in patients with metallosis and gross osteolysis.Competing Interest StatementJM is Chief Scientific Officer at Lyfstone. Lyfstone supplied calprotectin lateral flow tests free of charge for the study. IM is a paid consultant for Lyfstone.Funding StatementThis paper presents independent research funded by Orthopaedic Research UK (ORUK - grant number 526) the Biotechnology and Biological Sciences Research Council (BBSRC) Institute Strategic Programme Microbes in the Food Chain BB/R012504/1 and its constituent projects BBS/E/F/000PR10348 and BBS/E/F/000PR10349 (JOG).Author DeclarationsAll relevant ethical guidelines have been followed and any necessary IRB and/or ethics committee approvals have been obtained.YesAll necessary patient/participant consent has been obtained and the appropriate institutional forms have been archived.YesAny clinical trials involved have been registered with an ICMJE-approved registry such as ClinicalTrials.gov and the trial ID is included in the manuscript.Not ApplicableI have followed all appropriate research reporting guidelines and uploaded the relevant Equator, ICMJE or other checklist(s) as supplementary files, if applicable.YesAll data generated or analysed during this study are included in this published article (and its supplementary information files). ER -