PT - JOURNAL ARTICLE AU - Ruwanpathirana, Gihan P. AU - Williams, Robert C. AU - Masters, Colin L. AU - Rowe, Christopher C. AU - Johnston, Leigh A. AU - Davey, Catherine E. TI - Inter-scanner Aβ-amyloid PET harmonization using barrel phantom spatial resolution matching AID - 10.1101/2023.07.16.23292733 DP - 2023 Jan 01 TA - medRxiv PG - 2023.07.16.23292733 4099 - http://medrxiv.org/content/early/2023/07/18/2023.07.16.23292733.short 4100 - http://medrxiv.org/content/early/2023/07/18/2023.07.16.23292733.full AB - The standardized uptake value ratio (SUVR) is used to measure Aβ uptake in PET images of the brain. Variations in PET scanner technologies and image reconstruction techniques can lead to variability in images acquired from different scanners. This poses a challenge for Aβ-PET studies conducted across multiple centers. The aim of harmonization is to achieve consistent Aβ-PET measurements across scanners. In this study, the procedure of matching the spatial resolution of a barrel phantom measured in each PET scanner is proposed as a method of Aβ-PET harmonization, validated using subject data.Methods Three different PET scanners were used: the Siemens Biograph Vision 600, Siemens Biograph mCT, and Philips Gemini TF64. A total of five, eight, and five subjects were each scanned twice with [18F]-NAV4694 across Vision-mCT, mCT-Philips, and Vision-Philips scanner pairs. The Vision and mCT scans were reconstructed using various iterations, subsets, and post-reconstruction Gaussian smoothing, while one reconstruction configuration was used for the Philips scans. The full-width at half-maximum (FWHM) of each reconstruction configuration was calculated using [18F]-filled barrel phantom scans with the SNMMI phantom analysis toolkit. Regional SUVRs were calculated from 72 brain regions using the AAL3 atlas for each subject and reconstruction configuration. Statistical similarity between SUVRs was assessed using paired (within subject) t-tests for each pair of reconstructions across scanners; the higher the p-value, the greater the similarity between the SUVRs.Vision-mCT harmonization maximal statistical similarity (i.e., p-value) between regional SUVRs was achieved using a 4.10 mm FWHM Vision reconstruction with a 4.30 mm FWHM mCT reconstruction.Philips-mCT harmonization maximal statistical similarity between regional SUVRs was achieved using an 8.2 mm FWHM Philips reconstruction with a 9.35 mm FWHM mCT reconstruction.Philips-Vision harmonization a 9.1 mm FWHM Vision reconstruction had maximum statistical similarity with regional SUVRs from an 8.2mm FWHM Philips reconstruction. Reconstruction pairs that maximized statistical similarity, and supported a null hypothesis of being drawn from the same distribution, were selected as harmonised for each scanner pair.Conclusion Using data obtained from three sets of participants, each scanned on a different pair of PET scanners, using reconstruction configurations with matched barrel phantom spatial resolutions, we have demonstrated that Aβ-PET quantitation can be harmonised across scanners, producing SUVR values statistically likely to be drawn from the same distribution. This finding is encouraging for the use of different PET scanners in multi-centre trials, or updates during longitudinal studies.Competing Interest StatementThe authors have declared no competing interest.Funding StatementThis study was funded by the Australian Research Council and the National Health and Medical Research Council .Author DeclarationsI confirm all relevant ethical guidelines have been followed, and any necessary IRB and/or ethics committee approvals have been obtained.YesThe details of the IRB/oversight body that provided approval or exemption for the research described are given below:Ethics approval and consent to participate in this study were approved by the Austin Health Human Research Ethics Committee (HREC/18/Austin/201)I confirm that all necessary patient/participant consent has been obtained and the appropriate institutional forms have been archived, and that any patient/participant/sample identifiers included were not known to anyone (e.g., hospital staff, patients or participants themselves) outside the research group so cannot be used to identify individuals.YesI understand that all clinical trials and any other prospective interventional studies must be registered with an ICMJE-approved registry, such as ClinicalTrials.gov. I confirm that any such study reported in the manuscript has been registered and the trial registration ID is provided (note: if posting a prospective study registered retrospectively, please provide a statement in the trial ID field explaining why the study was not registered in advance).YesI have followed all appropriate research reporting guidelines, such as any relevant EQUATOR Network research reporting checklist(s) and other pertinent material, if applicable.YesAll data produced in the present study are available upon reasonable request to the authors