PT - JOURNAL ARTICLE AU - Csány, Gergely AU - Gergely, L. Hunor AU - Szalai, Klára AU - Lőrincz, Kende K. AU - Strobel, Lilla AU - Csabai, Domonkos AU - Hegedüs, István AU - Marosán-Vilimszky, Péter AU - Füzesi, Krisztián AU - Sárdy, Miklós AU - Gyöngy, Miklós TI - First clinical experience with a novel optical-ultrasound imaging device on various skin pathologies AID - 10.1101/2021.06.28.21259325 DP - 2021 Jan 01 TA - medRxiv PG - 2021.06.28.21259325 4099 - http://medrxiv.org/content/early/2021/07/02/2021.06.28.21259325.short 4100 - http://medrxiv.org/content/early/2021/07/02/2021.06.28.21259325.full AB - Objectives A compact handheld skin ultrasound imaging device has been developed that uses co-registered optical and ultrasound imaging to provide diagnostic information about the full skin depth and lesions encountered therein. The aim of the current work is to present the first clinical results of the device. Using additional photographic, dermoscopic and ultrasonic images as reference, the images from the device are assessed in terms of the detectability of the skin layer boundaries (between the epidermis, dermis, and subcutis), and in terms of image features produced by common skin lesions.Methods Combined optical-ultrasonic recordings of various types of common skin lesions (melanoma, basal cell carcinoma, keratosis, dermatofibroma, naevus, dermatitis, psoriasis) were taken with the device (N = 53) and compared with images from a reference portable skin ultrasound imager. The investigator and two additional independent experts evaluated and compared the images in terms of skin structure detectability and skin lesion features.Results Skin structure detectability was unanimously over 90 % for epidermis, dermis and lesion. Morphological and echogenicity information observed for melanoma, basal cell carcinoma, keratoses, dermatofibroma, naevi, atopic dermatitis, psoriasis were found consistent with those of the reference ultrasound device and relevant ultrasound images in the literature.Conclusions The presented device is able to obtain simultaneous in-vivo optical and ultrasound images of common skin lesions. This has the potential to provide relevant information in a number of settings to be investigated in the future, including preoperative planning of skin cancer treatment.Competing Interest StatementD. Cs., G. Cs., K. F., M. Gy., P. M-V., L. S. are employees of Dermus Kft. D. Cs., G. Cs., K. F., M. Gy., P. M-V., hold stock in Dermus Kft. L.H. G. is a contracted advisor of Dermus Kft.Funding StatementNo external funding was received.Author DeclarationsI confirm all relevant ethical guidelines have been followed, and any necessary IRB and/or ethics committee approvals have been obtained.YesThe details of the IRB/oversight body that provided approval or exemption for the research described are given below:The present study was performed under the ethical approval of the Hungarian National Institute of Pharmacy and Nutrition (approval number OGYEI/8317/2020), with all subjects providing their written informed consent.All necessary patient/participant consent has been obtained and the appropriate institutional forms have been archived.YesI understand that all clinical trials and any other prospective interventional studies must be registered with an ICMJE-approved registry, such as ClinicalTrials.gov. I confirm that any such study reported in the manuscript has been registered and the trial registration ID is provided (note: if posting a prospective study registered retrospectively, please provide a statement in the trial ID field explaining why the study was not registered in advance).YesI have followed all appropriate research reporting guidelines and uploaded the relevant EQUATOR Network research reporting checklist(s) and other pertinent material as supplementary files, if applicable.YesData are available on request from the authors.