TY - JOUR T1 - 7-Tesla Functional Cardiovascular MR Using Vectorcardiographic Triggering – overcoming the magnetohydrodynamic effect JF - medRxiv DO - 10.1101/2021.05.10.21256949 SP - 2021.05.10.21256949 AU - Christian Hamilton-Craig AU - Daniel Stäeb AU - Aiman Al Najjar AU - Kieran O’Brien AU - William Crawford AU - Sabine Fletcher AU - Markus Barth AU - Graham Galloway Y1 - 2021/01/01 UR - http://medrxiv.org/content/early/2021/05/14/2021.05.10.21256949.abstract N2 - Objective Ultra-high-field (B0 ≥ 7 Tesla) cardiovascular magnetic resonance (CMR) offers increased resolution. However, ECG gating is impacted by the magneto-hydrodynamic (MHD) effect distorting the ECG trace. We explored the technical feasibility of a 7T MR scanner using ECG trigger learning algorithm to quantitatively assess cardiac volumes and vascular flow.Methods 7T scans performed on 10 healthy volunteers on a whole-body research MRI (Siemens Healthcare, Erlangen, Germany) with 8 channel Tx/32 channel Rx cardiac coil (MRI Tools GmbH, Berlin, Germany). Vectorcardiogram ECG was performed using a learning phase outside of the magnetic field, with a trigger algorithm overcoming severe ECG signal distortions.Vectorcardiograms were quantitatively analyzed for false negative and false positive events. Cine CMR was performed after 3rd-order B0 shimming using a high-resolution breath-held ECG-retro-gated segmented spoiled gradient echo, and 2D phase contrast flow imaging. Artefacts were assessed using a semi-quantitative scale.Results 7T CMR scans were acquired in all patients (100%) using the VCG learning method. 3,142 R-waves were quantitatively analyzed, yielding sensitivity 97.6%, specificity 98.7%. Mean image quality score was 0.9, sufficient to quantitate both cardiac volumes, ejection fraction (EF), aortic and pulmonary blood flow. Mean LVEF was 56.4%, RVEF 51.4%.Conclusion Reliable cardiac ECG triggering is feasible in healthy volunteers at 7T utilizing a state-of-the-art 3-lead trigger device despite signal distortion from the MHD effect. This provides sufficient image quality for quantitative analysis. Other ultra-high-field imaging applications such as human brain functional MRI with physiologic noise correction may benefit from this method of ECG triggering.Key pointsUltra-high field 7 Tesla cardiac MRI is challenging due to the impact of the magneto-hydrodynamic (MHD) effect causing severe distortions in the ECG trace.Using VCG with a learning phase outside the ultra-high field magnet, the R waves can be adequately detected to perform high quality Cardiac MRI scans, overcoming signal distortion from the MHD effect.Competing Interest StatementThe authors have declared no competing interest.Funding StatementThis work was supported by UQ Academic Title Holder Research Fund and CAESIE: Connecting Australian European Science & Innovation Excellence Priming GrantAuthor DeclarationsI confirm all relevant ethical guidelines have been followed, and any necessary IRB and/or ethics committee approvals have been obtained.YesThe details of the IRB/oversight body that provided approval or exemption for the research described are given below:University of Queensland Ethics Review Board UQ approval 200500050All necessary patient/participant consent has been obtained and the appropriate institutional forms have been archived.YesI understand that all clinical trials and any other prospective interventional studies must be registered with an ICMJE-approved registry, such as ClinicalTrials.gov. I confirm that any such study reported in the manuscript has been registered and the trial registration ID is provided (note: if posting a prospective study registered retrospectively, please provide a statement in the trial ID field explaining why the study was not registered in advance).YesI have followed all appropriate research reporting guidelines and uploaded the relevant EQUATOR Network research reporting checklist(s) and other pertinent material as supplementary files, if applicable.YesData are available upon reasonable request(CMR)Cardiovascular magnetic resonance(MRI)Magnetic resonance imaging(SNR)Signal-to-noise ratio(MHD)Magneto-hydrodynamic(ECG)Electrocardiogam(VCG)Vectorcardiography(GRAPPA)GeneRalized Autocalibrating Partial Parallel Acquisition(SSFP)Steady state free precession(RF)Radiofrequency ER -