Abstract
Introduction Reliable, well-maintained laboratory equipment is essential for timely, accurate and reliable laboratory results for patient care, surveillance, and health system resilience. Tanzania implemented the Standard Medical Laboratory Equipment Guideline (SMLEG-2018) to harmonise procurement and improve efficiency across a tiered laboratory system and network. However, equipment functionality gaps remain a critical challenge for healthcare delivery. This study determined the functionality status of standardised laboratory analysers in Tanzania mainland and identified the drivers of non-functionality.
Methods We conducted a facility-based cross-sectional study to identify the causes of non-functionality of standardised chemistry, haematology, and urine analysers across mainland Tanzania. We assessed 3,044 analysers and used Pareto analysis to identify the primary drivers of non-functionality.
Results Of the 3,044 standardised analysers assessed, 987 (32.4%) chemistry 1,308 (43.0%) haematology, and 749 (24.6%) urine) assessed; 397 (13.0%) were non-functional. Proportionally, urine analysers were the most non-functional (n=176/749, 23.5%), followed by chemistry (n=135/987, 13.7%), and haematology (n=86/1308, 6.6%) analysers, p<0.001. Urine analysers had significantly higher non-functional equipment (23.5%), followed by Chemistry analysers (13.7%), than Haematology analysers (6.6%), p<0.001. Equipment breakdown was the leading cause of non-functionality, accounting for 70.9% (61/86), 70.5% (124/176) and 60.0% (81/135) of the non-functional haematology, urine and chemistry analysers, respectively, among all causes identified for each analyser type.
Conclusion Although the SMLEG 2018 implementation identified broad coverage of functional standardised laboratory equipment, persistent equipment breakdowns of more than 10% highlight a systemic, concerning gap that requires immediate corrective action to sustain high performance. Strengthening preventive maintenance, timely repair services and sustainable equipment lifecycle planning is critical to sustaining quality laboratory services for robust preventive and curative care.
Competing Interest Statement
The authors have declared that no competing interests exist.
Author Declarations
I confirm all relevant ethical guidelines have been followed, and any necessary IRB and/or ethics committee approvals have been obtained.
Yes
The details of the IRB/oversight body that provided approval or exemption for the research described are given below:
The study evaluated the implementation of SMLEG-2018 within the MoH framework for routine services without direct access to patient information therefore, MoH granted permission to publish the findings and a waiver through a letter (Ref. no. MP.76232/06), dated 19th May 2026. Before analysis, data was anonymised to ensure the confidentiality of facilities, equipment, and testing platforms.
I confirm that all necessary patient/participant consent has been obtained and the appropriate institutional forms have been archived, and that any patient/participant/sample identifiers included were not known to anyone (e.g., hospital staff, patients or participants themselves) outside the research group so cannot be used to identify individuals.
Yes
I understand that all clinical trials and any other prospective interventional studies must be registered with an ICMJE-approved registry, such as ClinicalTrials.gov. I confirm that any such study reported in the manuscript has been registered and the trial registration ID is provided (note: if posting a prospective study registered retrospectively, please provide a statement in the trial ID field explaining why the study was not registered in advance).
Yes
I have followed all appropriate research reporting guidelines, such as any relevant EQUATOR Network research reporting checklist(s) and other pertinent material, if applicable.
Yes
Data Availability
All data are in the manuscript and the supporting information files