Abstract
Background Indwelling urinary catheters can cause pain, urethral irritation and trauma and provide surfaces for microbial colonisation and biofilm formation. The LubriShield® Foley catheter incorporates a covalently bonded, superhydrophilic antifouling hydrogel designed to reduce surface friction and limit biofilm accumulation without releasing an antimicrobial agent. This study evaluated its safety and clinical performance compared with an uncoated silicone Foley catheter, including usability and patient comfort, and explored catheter-associated microbial colonisation and biofilm formation.
Methods This prospective, randomised, single-blind, single-centre controlled clinical investigation enrolled 33 adults requiring indwelling urinary catheterisation for 72 hours to 14 days. Participants received either the LubriShield® Foley catheter (n = 16) or an uncoated Sterimed all-silicone Foley catheter (n = 17). The primary endpoint was safety based on adverse-event reporting; the prespecified safety criterion was the absence of a serious adverse device effect in the LubriShield® group. Secondary endpoints included investigator-rated ease and functionality of catheter insertion and removal and participant-reported pain and irritation/discomfort assessed using 0–10 numerical rating scales. Urine microbiology, catheter colonisation and catheter-level biofilm measurements were exploratory endpoints. Analyses were performed in the as-treated population. Numerical rating scale outcomes were compared using two-sided Mann–Whitney U tests; exploratory biofilm analyses used permutation tests with bootstrap confidence intervals.
Results No serious adverse events or serious adverse device effects occurred. Nine non-serious adverse events were reported and resolved completely. At least one adverse event occurred in 4/16 (25.0%) participants in the LubriShield® group and 4/17 (23.5%) participants in the control group. The prespecified safety criterion, defined as the absence of a SADE in the LubriShield® group, was met. LubriShield® was rated more favourably for ease of insertion (p = 0.0096), functionality during insertion (p = 0.0051), ease of removal (p = 0.0037) and functionality during removal (p = 0.0085). Participants reported significantly less pain and irritation/discomfort during insertion, while the catheter was indwelling and during removal (all p < 0.05). Outcomes 7 ± 1 days after catheter removal were similar between groups. Absolute biofilm biomass and bacterial and fungal loads did not differ significantly. In a post hoc exploratory normalised analysis, the biofilm-to-bacterial-load ratio was lower with LubriShield® (median 0.052 vs 0.080; permutation p = 0.043), although the bootstrap 95% confidence interval for the between-group difference included zero.
Conclusions The LubriShield® Foley catheter demonstrated a favourable short-term safety profile and was associated with improved investigator-rated usability and participant-reported comfort compared with an uncoated silicone Foley catheter. A post hoc exploratory analysis was compatible with reduced biofilm accumulation relative to bacterial load, but this finding is hypothesis-generating only and requires confirmation in larger, adequately powered studies with prespecified biofilm endpoints.
Trial registration Clinical Trials Registry–India (CTRI), CTRI/2025/05/086528, prospectively registered on 7 May 2025, before participant enrolment. Subsequently registered with ClinicalTrials.gov, NCT07070921.
Competing Interest Statement
AIR serves as a consultant to CytaCoat AB. SS, PW and LB are employees of CytaCoat AB, and GB is its CEO. KL and JA are affiliated with Segulah Medical Acceleration, an investor and shareholder in CytaCoat AB. JA has also received consulting fees from CytaCoat AB. CytaCoat AB is the sponsor of the study and manufacturer of the investigational LubriShield® Foley catheter. SM, MK and RS declare no competing interests.
Clinical Trial
NCT07070921
Author Declarations
I confirm all relevant ethical guidelines have been followed, and any necessary IRB and/or ethics committee approvals have been obtained.
Yes
The details of the IRB/oversight body that provided approval or exemption for the research described are given below:
The Institutional Ethics Committee of St. John's Medical College Hospital, Bengaluru, India, gave ethical approval for this work.
I confirm that all necessary patient/participant consent has been obtained and the appropriate institutional forms have been archived, and that any patient/participant/sample identifiers included were not known to anyone (e.g., hospital staff, patients or participants themselves) outside the research group so cannot be used to identify individuals.
Yes
I understand that all clinical trials and any other prospective interventional studies must be registered with an ICMJE-approved registry, such as ClinicalTrials.gov. I confirm that any such study reported in the manuscript has been registered and the trial registration ID is provided (note: if posting a prospective study registered retrospectively, please provide a statement in the trial ID field explaining why the study was not registered in advance).
Yes
I have followed all appropriate research reporting guidelines, such as any relevant EQUATOR Network research reporting checklist(s) and other pertinent material, if applicable.
Yes
Data Availability
The datasets generated and/or analysed during the current study are not publicly available due to participant confidentiality and data protection requirements but may be available from the corresponding author on reasonable request, subject to applicable ethical, legal and data protection requirements.
List of abbreviations
- AE
- adverse event
- APACHE II
- Acute Physiology and Chronic Health Evaluation II
- CAUTI
- catheter-associated urinary tract infection
- CFU
- colony-forming units
- CI
- confidence interval
- COPD
- chronic obstructive pulmonary disease
- CTRI
- Clinical Trials Registry–India
- IEC
- Institutional Ethics Committee
- IQR
- interquartile range
- LB
- Luria–Bertani
- NRS
- numerical rating scale
- OD₅₉₅
- optical density at 595 nm
- PBS
- phosphate-buffered saline
- RTA
- road traffic accident
- SADE
- serious adverse device effect
- SAE
- serious adverse event
- SD
- standard deviation
- SDA
- Sabouraud dextrose agar.