Abstract
Background Artificial intelligence computer-aided diagnosis (AI CAD) algorithms for screening mammography have shown promise, but independent head-to-head comparisons of commercial algorithms on large, diverse cohorts remain limited.
Methods 786,124 mammography screening exam performed from January 1 2022 to December 31 2023 were identified from electronic medical records. Three FDA-cleared commercial algorithms (A, B, D) and the open-source academic model Mirai (C), were applied to the four standard screening views. Breast cancer within 12 months was ascertained through cancer registry linkage. Algorithms were compared by AUROC; sensitivity, specificity, and PPV at matched AI-positive rates of 5%, 10%, 15%, 20%, 25%; and the proportion of false negatives classified as AI-positive.
Results Of 707,922 examinations with complete data, 4,436 were associated with cancer. Radiologists recalled 6.8% of examinations, with a sensitivity of 67.2%. Algorithm D had the highest AUROC (0.849), followed by algorithm B (0.818), Mirai (0.817), and algorithm A (0.815). Algorithm D also had the highest sensitivity at every AI-positive rate, from 55.3% (95% CI: 54.1, 56.6) at 5% to 77.9% (95% CI: 76.9, 79.0) at 25%, with CIs that did not overlap those of the other algorithms. No algorithm matched radiologist sensitivity at AI-positive rates of 10% or less. Algorithm B flagged a proportion of radiologist false negatives similar to that of algorithm D from 5% through 20% (20.8% vs 20.5% at 5%), despite its lower standalone sensitivity.
Conclusions AI CAD algorithms differed meaningfully in performance on the same large, diverse screening examination cohort, and an open-source academic model performed comparably to two of the three commercial products. Independent local validation at clinically relevant operating points should precede adoption of mammography AI, and algorithm choice and threshold should be matched to the intended workflow.
Competing Interest Statement
The authors have declared no competing interest.
Author Declarations
I confirm all relevant ethical guidelines have been followed, and any necessary IRB and/or ethics committee approvals have been obtained.
Yes
The details of the IRB/oversight body that provided approval or exemption for the research described are given below:
The IRB of Kaiser Permanente gave ethical approval for this work
I confirm that all necessary patient/participant consent has been obtained and the appropriate institutional forms have been archived, and that any patient/participant/sample identifiers included were not known to anyone (e.g., hospital staff, patients or participants themselves) outside the research group so cannot be used to identify individuals.
Yes
I understand that all clinical trials and any other prospective interventional studies must be registered with an ICMJE-approved registry, such as ClinicalTrials.gov. I confirm that any such study reported in the manuscript has been registered and the trial registration ID is provided (note: if posting a prospective study registered retrospectively, please provide a statement in the trial ID field explaining why the study was not registered in advance).
Yes
I have followed all appropriate research reporting guidelines, such as any relevant EQUATOR Network research reporting checklist(s) and other pertinent material, if applicable.
Yes
Footnotes
Funding: Kaiser Permanente Division of Research Delivery and Applied Research Program
Conflicts of interest: None
Data Availability
All data produced in the present study are available upon reasonable request to the authors