Abstract
Background Cardiorespiratory fitness and muscle strength are valid markers of health and strong predictors of mortality and morbidity. The tests used to measure these variables require in-person visits with specialized equipment and trained personnel – leading to organizational constraints both for patients and hospitals, and making them difficult to implement at a large scale. In this context, technologies embedded in smartphones offer new opportunities to develop remote tests.
Objectives This study aimed to test the validity and reliability of MediEval, a newly developed app-based medical device that allows individuals to perform the 6-minute walk test (6MWT) and the 30-second sit-to-stand (30s-STS) test on their own using GPS signal and camera detection with a skeleton extraction algorithm.
Methods A total of 53 healthy adults performed the two tests in three different sessions to determine the intra- and inter-day reproducibility. Test validity was assessed by comparing the results obtained from the app to gold standard measures. Pearson correlations and concordance correlation coefficients, the relative measurement error, intraclass correlation coefficients, the standard error of measure and the minimal detectable change were computed for each test.
Results The results revealed high to excellent validity of the app in comparison to gold standards (ρ=0.84 for the 6MWT and ρ=0.99 for the 30s-STS test) with low relative measurement error. The mean differences between the app and the gold standard measures were 8.96m for the 6MWT and 0.28 repetition for the 30s-STS test. Both tests had good test-retest reliability (ICCs = 0.77). The minimal detectable changes were respectively 97.56 meters for the 6MWT and 7.37 repetitions for the 30s-STS test.
Discussion The MediEval medical device proposes valid and reproducible measures of the 6MWT and the 30s-STS test. This device holds promise for monitoring the physical activity of large epidemiologic cohorts while refining patient experience and improving the scalability of the healthcare system. Considering minimal detectable change values, it may be important to ask participants to perform several tests and average them to improve accuracy. Future studies in clinical context are needed to evaluate the responsiveness and the smallest detectable changes of the device for specific populations with chronic diseases.
Data, code & supplemental materials https://osf.io/4hzke/
Competing Interest Statement
AC and MD declare that they have no competing interests. AM PhD grant is funded by the French National Association for Research and Technology (ANRT) and Kiplin. MB is employed by Kiplin.
Funding Statement
The work of AM is supported by an ANRT grant (Cifre PhD Thesis) and by the company Kiplin. The funders had no input in the design of the study and no influence on the interpretation or publication of the study results.
Author Declarations
I confirm all relevant ethical guidelines have been followed, and any necessary IRB and/or ethics committee approvals have been obtained.
Yes
The details of the IRB/oversight body that provided approval or exemption for the research described are given below:
The study protocol was approved by the local ethics committee: the CERSTAPS (which evaluates research protocols submitted by sport and exercise sciences teams at national level) gave ethical approval for this work (IRB00012476- 2022-26-04-177).
I confirm that all necessary patient/participant consent has been obtained and the appropriate institutional forms have been archived, and that any patient/participant/sample identifiers included were not known to anyone (e.g., hospital staff, patients or participants themselves) outside the research group so cannot be used to identify individuals.
Yes
I understand that all clinical trials and any other prospective interventional studies must be registered with an ICMJE-approved registry, such as ClinicalTrials.gov. I confirm that any such study reported in the manuscript has been registered and the trial registration ID is provided (note: if posting a prospective study registered retrospectively, please provide a statement in the trial ID field explaining why the study was not registered in advance).
Yes
I have followed all appropriate research reporting guidelines, such as any relevant EQUATOR Network research reporting checklist(s) and other pertinent material, if applicable.
Yes
Data Availability
The anonymized data used in this study and the R code are available on the Open Science Framework (https://osf.io/4hzke/).