ABSTRACT
Background Many research studies seek to improve vital sign monitoring to enhance the conditions under which doctors and caregivers track patients ‘ health. Non-invasive and contactless monitoring has emerged as an optimal solution for this problem, with telemedicine, self-monitoring, and wellbeing tools being the next generation of technology in the biomedical field. However, there is worldwide concern about the general purpose and bias towards a certain demographic group of these techniques. In particular, skin tone and the accuracy of monitoring dark skin tone groups have been key questions among researchers, with the lack of results and studies contributing to this uncertainty.
Methods This paper proposes a benchmark for remote monitoring solution against a medical device across different skin tone people. Around 330 videos from 90 different patients were analyzed, and Heart Rate and Heart Rate Variability were compared across different subgroups. The Fitzpatrick scale (1-6) was used to classify participants into three skin tone groups: 1 and 2; 3 and 4; 5 and 6.
Results The results showed that our proposed methodology was able to estimate heart rate with a mean absolute error of 3 bpm across all samples and subgroups. Moreover, for Heart Rate Variability (HRV) metrics, we achieved the following results, in terms of Mobility Assistive Equipment (MAE): HRV-IBI (Inter-Beat-Interval) of 10 ms; HRV-SDNN (Standard Deviation of Normal to Normal heartbeats) of 14 ms and HRV-RMSSD (Root Mean Square of Successive Differences between normal heartbeats) of 22 ms. No significant performance decrease was found for any skin tone group, and there was no error trend towards a certain group.
Conclusions The study showed that our methodology meets acceptable agreement levels for the proposed metrics and is well-suited for users who want to understand their general health and wellness. Furthermore, the experiments showed that skin tone had no impact on the results, which remained within the same range across all groups.
Competing Interest Statement
The authors have declared no competing interest.
Funding Statement
None
Author Declarations
I confirm all relevant ethical guidelines have been followed, and any necessary IRB and/or ethics committee approvals have been obtained.
Yes
I confirm that all necessary patient/participant consent has been obtained and the appropriate institutional forms have been archived, and that any patient/participant/sample identifiers included were not known to anyone (e.g., hospital staff, patients or participants themselves) outside the research group so cannot be used to identify individuals.
Yes
I understand that all clinical trials and any other prospective interventional studies must be registered with an ICMJE-approved registry, such as ClinicalTrials.gov. I confirm that any such study reported in the manuscript has been registered and the trial registration ID is provided (note: if posting a prospective study registered retrospectively, please provide a statement in the trial ID field explaining why the study was not registered in advance).
Yes
I have followed all appropriate research reporting guidelines, such as any relevant EQUATOR Network research reporting checklist(s) and other pertinent material, if applicable.
Yes
Data Availability
Researchers interested in access to the data may contact Nikhil Sehgal at nikhil.sehgal{at}vastmindz.com, also see https://vastmindz.com/contact/. The authors will assist with any negotiable data use agreements and gain access to the data for any replication efforts following publication.
Abbreviations
- AI
- Artificial Intelligence
- HR
- Heart Rate
- HRV
- Heart Rate Variability
- IBI
- Inter-Beat-Interval
- PPG
- Photoplethysmography
- RMSSD
- Root Mean Square of Successive Differences between normal heartbeats SDNN Standard Deviation of Normal to Normal heartbeats