Background The gold standard for diagnosis of Lisfranc instability is direct visualization in the operation room while the examination techniques is still unstandardized and non-reproducible. We aimed to introduce a novel reproducible intraoperative mechanical testing method (Listract test) for intraoperative isolated Lisfranc instability assessment.
Methods The Lisfranc ligament between the first cuneiform (C1) and second metatarsus (M2) in eight lower leg cadaveric specimens were dissected to replicate C1-M2 Lisfranc instability by eight foot and ankle surgeons. The 50N distraction force was applied in the direction of the C1-M2 ligament. Three methods of fixation - flexible fixation, metal screw, and bio-absorbable screw- were used to fix the injury, and Listract test was applied again after fixation. Besides intraoperative assessment, C1-M2 diastasis and area were measured using radiographs for assessment of Lisfranc instability.
Results The sensitivity and specificity of the Listract test for detection of C1-M2 instability were 100% and 100% intraoperatively, 33.3% and 95.2% using radiographic diastasis measurement, and 63.2% and 38% using area measurement, respectively. The Listract test had a specificity and sensitivity of 100% and 96% for intraoperative assessment, 87.5% and 64.3 for radiographic C1-M2 diastasis, and 48% and 50% for radiographic area.
Conclusion The Listract test is a simple, reproducible, and replicable intraoperative method for evaluating the Lisfranc joint for instability. Developing a device with this mechanism can help clinicians confirm the diagnosis and provide appropriate treatment particularly for equivocal diagnoses.
Authors’ contributions SAE: Conception, design, and conducting the study, analysis, drafting the manuscript; VM, DOH, SHSH, FR: conducting the study, analysis, preparing the manuscript draft; LBP, JK, GW, PK: conducting the study, manuscript preparation, discussion.
Competing Interest Statement
The authors have declared no competing interest.
Funding Statement
Acknowledgement This study was logistically and financially sponsored by FARIL-MGH Center, Ossio Inc., and Depuy Synthes, who helped us provide the instruments and conduct the experiment. Conflict of Interest None of the sponsoring companies have participated, interfered, or were aware of the outcomes of the study until the manuscript was publicly published. None of the authors has any conflict of interest with the sponsoring companies.
Author Declarations
I confirm all relevant ethical guidelines have been followed, and any necessary IRB and/or ethics committee approvals have been obtained.
Yes
I confirm that all necessary patient/participant consent has been obtained and the appropriate institutional forms have been archived, and that any patient/participant/sample identifiers included were not known to anyone (e.g., hospital staff, patients or participants themselves) outside the research group so cannot be used to identify individuals.
Yes
I understand that all clinical trials and any other prospective interventional studies must be registered with an ICMJE-approved registry, such as ClinicalTrials.gov. I confirm that any such study reported in the manuscript has been registered and the trial registration ID is provided (note: if posting a prospective study registered retrospectively, please provide a statement in the trial ID field explaining why the study was not registered in advance).
Yes
I have followed all appropriate research reporting guidelines and uploaded the relevant EQUATOR Network research reporting checklist(s) and other pertinent material as supplementary files, if applicable.
Yes
Data Availability
All data produced in the present study are available upon reasonable request to the authors.