Abstract
Background Bioelectrical impedance analysis (BIA) is gaining popularity as a body composition assessment tool for patients. Although it has been studied and validated in different populations, age groups, and clinical settings, including critically ill patients, there are concerns about BIA reproducibility and reliability in different device types and postures. This study aimed to evaluate the reliability of BIA according to different devices, postures, and lead types.
Methods Cross-sectional observational data of 74 healthy volunteers (32 women and 42 men) were collected. We used two types of devices, three types of postures (standing, sitting, and lying), and two lead types (clamp lead and adhesive lead) to measure the whole-body phase angle (phA) at a single frequency of 50 kHz. The measurements were validated using the intraclass correlation coefficient (ICC) and Bland–Altman plot analysis.
Results All phA measurements recorded with two types of devices, three different postures, and two types of leads were equivalent to each other (mean ICC = 0.9932, 95% confidence interval (CI) 0.9905–0.0053). The average mean difference in phA was 0.31 (95% CI 0.16– 0.46). The largest phA value was measured using BWA with an adhesive-type lead in the lying position. There were no differences between standing and sitting positions.
Conclusion This is the first study to show the consistency and reliability of BIA in measuring phA using different devices, lead types, and postures. This could provide the confidence that BIA can be used in various clinical settings.
Competing Interest Statement
The authors have declared no competing interest.
Funding Statement
The funders had no role in study design, data collection and analysis, decision to publish, or preparation of the manuscript.
Author Declarations
I confirm all relevant ethical guidelines have been followed, and any necessary IRB and/or ethics committee approvals have been obtained.
Not Applicable
The details of the IRB/oversight body that provided approval or exemption for the research described are given below:
This study was approved by the Korea University Institutional Review Board (IRB No. 2020AN0145)
I confirm that all necessary patient/participant consent has been obtained and the appropriate institutional forms have been archived, and that any patient/participant/sample identifiers included were not known to anyone (e.g., hospital staff, patients or participants themselves) outside the research group so cannot be used to identify individuals.
Not Applicable
I understand that all clinical trials and any other prospective interventional studies must be registered with an ICMJE-approved registry, such as ClinicalTrials.gov. I confirm that any such study reported in the manuscript has been registered and the trial registration ID is provided (note: if posting a prospective study registered retrospectively, please provide a statement in the trial ID field explaining why the study was not registered in advance).
Not Applicable
I have followed all appropriate research reporting guidelines and uploaded the relevant EQUATOR Network research reporting checklist(s) and other pertinent material as supplementary files, if applicable.
Not Applicable
Data Availability
Data cannot be shared publicly because of IRB regulation.