Abstract
Background The usefulness of opportunistic arrhythmia screening strategies, using an electrocardiogram (ECG) or other methods for random “snapshot” assessments is limited by the unexpected and occasional nature of arrhythmias, leading to a high rate of missed-diagnosis.
We have previously validated a cardiac monitoring system for AF detection pairing simple consumer-grade Bluetooth low-energy (BLE) heart rate (HR) sensors with a smartphone application (RITMIAtm, Heart Sentinel srl, Italy). In the current study we test a significant upgrade to the abovementioned system, thanks to the technical capability of new HR sensors to run algorithms on the sensor itself and to acquire (and store on-board) single-lead ECG strips, if asked to do so.
Methods and Result We have reprogrammed a HR monitor intended for sports use (Movensense HR+) to run our proprietary RITMIA algorithm code in real-time, based on RR analysis, so that if any type of arrhythmia is detected it triggers a brief retrospective recording of a single-lead ECG, providing tracings of the specific arrhythmia for later consultation.
We report the initial data on the behavior, feasibility and high diagnostic accuracy of this ultra-low weight customized device for standalone automatic arrhythmia detection and ECG recording, when several types of arrhythmias were simulated, under different baseline conditions.
Conclusions The customized device was capable to detect all types of simulated arrhythmias and correctly triggered an visually interpretable ECG tracing. Future human studies are needed to address real-life accuracy of this device.
Competing Interest Statement
Some of the authors (NG, CR, BDM) are part-time collaborators of the startup enterprise owning the intellectual property of the algorithms, Heartsentinel srl, Italy. NG and CR are the inventors of the RITMIA algorithm used by the tested device.
Funding Statement
This study did not receive any funding
Author Declarations
I confirm all relevant ethical guidelines have been followed, and any necessary IRB and/or ethics committee approvals have been obtained.
Yes
I confirm that all necessary patient/participant consent has been obtained and the appropriate institutional forms have been archived, and that any patient/participant/sample identifiers included were not known to anyone (e.g., hospital staff, patients or participants themselves) outside the research group so cannot be used to identify individuals.
Yes
I understand that all clinical trials and any other prospective interventional studies must be registered with an ICMJE-approved registry, such as ClinicalTrials.gov. I confirm that any such study reported in the manuscript has been registered and the trial registration ID is provided (note: if posting a prospective study registered retrospectively, please provide a statement in the trial ID field explaining why the study was not registered in advance).
Yes
I have followed all appropriate research reporting guidelines and uploaded the relevant EQUATOR Network research reporting checklist(s) and other pertinent material as supplementary files, if applicable.
Yes
Data Availability
All data produced in the present work are contained in the manuscript.