Abstract
Purpose To implement a tool for real time image quality feedback for chest radiographs into the clinical routine and to evaluate the effect of the system on the image quality of the acquired radiographs.
Materials and Methods A real time Artificial Intelligence (AI) image quality feedback tool is developed that analyzes chest PA x-rays right after the completion of the examination at the x-ray system and provides visual feedback to the system operator with respect to adherence to desired standards of collimation, patient rotation and inspiration. In order to track image quality changes over time, results were compared to image quality assessment for images, acquired prior to system implementation.
Results Compared to the image quality prior to the installation of the real time image quality feedback solution, it is shown that a relative increase of images with optimal image quality with respect to collimation, patient rotation and inspiration is achieved by 30% (p<0.01). A relative improvement of 28% (p<0.01) is observed for the increase of images with optimal collimation, followed by a relative increase of 4% (p<0.01) of images with optimal inspiration. Finally, a detailed analysis is presented that shows that the average unnecessarily exposed area is reduced by 34% (p<0.01).
Discussion This study shows that it is possible to significantly improve image quality using a real time AI-based image quality feedback tool. The developed tool not only provides objective and impartial criticism and helps x-ray operators identify areas for improvement, but also gives positive feedback.
Key Findings
A substantial amount of images acquired in the clinical routine does not suffice the international guidelines
Continuous AI-based image quality feedback to the x-ray system operator in the clinical routine leads to a significant image quality improvement over time
Using the developed tool, the overall fraction of images with optimal patient positioning could be improved by 30%, followed by a 34% decrease of unnecessarily exposed area.
Competing Interest Statement
A.Y. and N.W. are employees of Philips Medical Systems DMC GmbH, T.H. is employee of Philips Research.
Funding Statement
There was no funding for this work
Author Declarations
I confirm all relevant ethical guidelines have been followed, and any necessary IRB and/or ethics committee approvals have been obtained.
Yes
The details of the IRB/oversight body that provided approval or exemption for the research described are given below:
Ethical Board of the University Witten/Herdecke
All necessary patient/participant consent has been obtained and the appropriate institutional forms have been archived.
Yes
I understand that all clinical trials and any other prospective interventional studies must be registered with an ICMJE-approved registry, such as ClinicalTrials.gov. I confirm that any such study reported in the manuscript has been registered and the trial registration ID is provided (note: if posting a prospective study registered retrospectively, please provide a statement in the trial ID field explaining why the study was not registered in advance).
Yes
I have followed all appropriate research reporting guidelines and uploaded the relevant EQUATOR Network research reporting checklist(s) and other pertinent material as supplementary files, if applicable.
Yes
Data Availability
Image quality statistics obtained during this study are available upon legitimate request