Abstract
Background Bubble CPAP may be used in infants with suspected or confirmed COVID-19. Electrostatic filters may reduce cross-infection. This study aims to determine if including a filter in the bubble CPAP circuit impacts stability of pressure delivery.
Methods A new electrostatic filter was placed before (pre) or after (post) the bubble CPAP generator, or with no filter (control) in an in vitro study. Pressure was recorded at the nasal interface for 18 h (6 L/min; 7 cmH2O) on three occasions for each configuration. Filter failure was defined as pressure >9 cmH20 for 60 continuous minutes. The filter was weighed before and after each experiment.
Results Mean (SD) time to reach the fail-point was 257 (116) min and 525 (566) min for filter placement pre- and post-CPAP generator, respectively. Mean pressure was higher throughout in the pre-generator position compared to control. The filter weight was heavier at study end in the pre-compared to the post-generator position.
Conclusions Placement of the filter at the pre-generator position in a bubble CPAP circuit should be avoided due to unstable mean pressure. Filters are likely to become saturated with water over time. The post-generator position may accommodate a filter, but regular pressure monitoring and early replacement are required.
Competing Interest Statement
The authors have declared no competing interest.
Funding Statement
No specific funding was received for this study
Author Declarations
I confirm all relevant ethical guidelines have been followed, and any necessary IRB and/or ethics committee approvals have been obtained.
Yes
The details of the IRB/oversight body that provided approval or exemption for the research described are given below:
This study was a bench top study and so no specific approval was required.
All necessary patient/participant consent has been obtained and the appropriate institutional forms have been archived.
Yes
I understand that all clinical trials and any other prospective interventional studies must be registered with an ICMJE-approved registry, such as ClinicalTrials.gov. I confirm that any such study reported in the manuscript has been registered and the trial registration ID is provided (note: if posting a prospective study registered retrospectively, please provide a statement in the trial ID field explaining why the study was not registered in advance).
Yes
I have followed all appropriate research reporting guidelines and uploaded the relevant EQUATOR Network research reporting checklist(s) and other pertinent material as supplementary files, if applicable.
Yes
Data Availability
Data is available upon request